When a red LED appears, try to detect the cause of the fault, and eliminate it as soon as
possible.
Red LED calculator
Possible causes
Action
Loss of communication on the CAN bus
Switch to maintenance mode and reboot the
calculator.
CAN bus power shorted
Switch to maintenance mode and reboot the
calculator.
Calculator probe disconnected from the RFID
card
Connect the temperature probe.
Ethernet network disconnected
Disconnect and connect the Ethernet cable.
Insufficient SD Card memory space
Verify that a notification has been sent to the
third-party system. Contact the administrator.
Power failure of at least one drawer
Verify that a notification has been sent to the
third-party system. Contact the administrator.
Battery charger Failure
Verify that a notification has been sent to the
third-party system. Contact the administrator.
Contact the supplier of your device in case of malfunction. Do not repair or modify the device
without prior authorization from Biolog-Id. Any maintenance operation must be preceded by
an SST-A30 mode change to enter maintenance mode (see GUI Manual).
7
Warranty
Any non-compliance with the recommendations will result in a breach of the warranty.
8
Transport
Upon receipt of the SST-A30, check that it has not been damaged during transport. If you
notice any transport damage, immediately contact the carrier or your dealer with the delivery
note or the purchase order.
9
Manufacturer Responsibility
The manufacturer's responsibility will not be held liable in case of:
•
Non-compliance with manufacturer's recommendations on the installation.
•
Intervention or repairs made by unauthorized persons by the manufacturer.
•
Use on an electrical installation that does not conform to the regulations in force
regulations.
•
The uses other than those specified in this manual.
•
Use of accessories (RFID chip temperature sensor ...) other than those provided by
Biolog-Id.