User Manual Smart Storage Agitator 30
Version
: 1.0
Validation date of the manual: 01/08/2021
25 of 29
5
Cleaning Instructions
This chapter explains the procedure for cleaning the SST-A30.
For proper operation, clean the SST-A30 at least once a month and more if necessary.
Only personnel qualified by the establishment are authorized to clean the SST-A30. Cleaning
personnel should be aware of the operation of the SST-A30 and its documentation, especially
the safety instructions.
The cleaning must be done as follows:
✓
Put the SST-A30 in maintenance mode.
✓
Move the bags into another agitator.
✓
Use a chemically compatible spray product with SST-A30 component materials,
combining cleaning and disinfecting and scrubbing with a soft cloth.
1. Apply the disinfectant detergent spray to the
area to be treated or to a nonwoven wipe.
2. Distribute the product.
To clean and maintain the proper functioning of the SST-A30, we advise you to follow the
instructions below.
Before each cleaning, make sure to put the SST-A30 in maintenance mode (see GUI
manual).
Risks of material damage caused by the use of unsuitable cleaning utensils, high-
pressure cleaner, or water spray or under pressure.
Do not use cleaning products containing:
- Acids and halogenated compounds (chlorides, bromides, iodides)
- Strongly acidic salts, e.g., descaler based on formic acid and amino sulfonic acid.
- Pipeline stripper, hydrochloric acid, silver cleaner.
- Chlorine.
- Abrasive and scouring components (scouring powder, steel wool)
- Polishing products, waxes, bleaching agents.
It is imperative to observe the instructions of the manufacturer of the cleaning agent
used with regard to temperature, dosage, time of action, etc.
After all the cleaning operations: verify that the unit is operational.