User Manual Smart Storage Agitator 30
Version
: 1.0
Validation date of the manual: 01/08/2021
22 of 29
Electromagnetic immunity, radio frequencies
The SST-A30 is intended to be used in the magnetic and electromagnetic environment
described in the table below. The user and the installer must guarantee the compliance of the
electromagnetic environment.
Immunity test
Test level
Compliance level
Electromagnetic
environment/comments
WARNING: Portable RF communication devices (including peripherals such as antenna cables and
external antennas) should not be used within 30 cm (12 inches) of any part of the PRD_7150200A,
including cables specified by the manufacturer. Otherwise, the performance of these devices could
be adversely affected.
Radiated RF
electromagnetic fields
(IEC61000-4-3)
3 V/m
80 MHz to 2.7 GHz
80% MA at 1 kHz
3 V/m
80 MHz to 2.7 GHz
80% MA at 1 kHz
Professional healthcare
facility environment.
Proximity fields
emitted by wireless
RF communication
devices (IEC 61000-4-
3 interim
method)
9 V/m
710 MHz, 745 MHz,
780 MHZ, 5240 MHz,
5550 MHz, 5785 MHz
27 V/m
385 MHz
28 V/m
450 MHz, 810 MHz,
870 MHz, 930 MHz,
1720 MHz, 1845
MHz,
1970 MHz, 2450 MHz
9 V/m
710 MHz, 745 MHz,
780 MHZ, 5240 MHz,
5550 MHz, 5785 MHz
27 V/m
385 MHz
28 V/m
450 MHz, 810 MHz,
870 MHz, 930 MHz,
1720 MHz, 1845 MHz,
1970 MHz, 2450 MHz
Professional healthcare
facility environment.
Conducted
disturbances, induced
by RF
fields (IEC610004-6)
3 V
150 KHz to 80 MHz
6 V within ISM band
and
band ranging from
0.15 MHZ to 80 MHZ,
80% MA at 1 KHz
3 V
150 KHz to 80 MHz
6 V within ISM band
and
band ranging from
0.15 MHZ to 80 MHZ,
80% MA at 1 KHz
Professional healthcare
facility environment.