Veterinary monitor user’s manual
Chapter 4-15
Maintenance
Click and open the dialog of system maintenance
Factory
Manufacturer maintenance is not an operation option for users and it must be
operated by the technical and maintenance personnel authorized by manufacturer.
CO
2
Gain Cal
Conduct gain calibration on the sidestream CO
2
module. This function is
only valid on sidestream CO
2
and when the sampling pump has been started.
CO
2
Cal Mode
Open or close the CO
2
calibration mode. When conducting calibration on
sidestream CO
2
, set the CO2 cal mode to ON.
HUM
Select the frequency of the AC power supply and options are
50Hz
and
60Hz
.
It is mainly configured according to the frequency of local power supply.
Gas zero
Conduct zero calibration on mainstream CO
2
module or anesthesia gas module.
Press this button, the following dialog will pop up. Select
〈
OK
〉
to conduct zero-calibration
operation. If
〈
Cancel
〉
is selected, the zero-calibration will not be implemented.
Note: The zero-calibration of Gas is only valid on the mainstream CO
2
module
and AG module of IRMA Company.