Veterinary monitor user’s manual
Chapter 1-13
Total cycle time
20 to 45s typical (dependent on heart rate and
motion artifact)
Overpressure Protection
Horse
Dog
Cat
Hardware and software double protections
315±10 mmHg
265±10 mmHg
255±10 mmHg
Intervals for AUTO
measurement time
1,2,3,4,5,10,15,20,30,45,60,90 minutes
2,4,8 hours
Range of alarm
SYS
0~300 mmHg, continuously adjustable
between upper limit and lower limit
DIA
0~300 mmHg, continuously adjustable
between upper limit and lower limit
MEAN
0~300 mmHg, continuously adjustable
between upper limit and lower limit
Alarm indication
Sound and light alarming
Measurement Mode
Horse
Manual, Auto and STAT
Dog
Manual, Auto and STAT
Cat
Manual, Auto
SpO2
BLT-SpO
2
Measurement Range
0~100%
Resolution
1%
Accuracy
At 70~100%, ±2%
At 0~69%, unspecified
Data update period
<13s
Alarm
User-selectable upper and lower SpO
2
limits
PR
Measurement Range
25~250 bpm
Resolution
1 bpm
Accuracy
±1% or ±1 bpm, whichever is greater
Data update period
<13s
Alarm
User-selectable upper and lower pulse rate limits
Nellcor-SpO
2
( option)
Measurement Range
0~100%