Veterinary Monitor User’s Manual

BIOLIGHT MEDITECH USA

535 Broadhollow Road Ste B-24, Melville, New York 747

Http//www.biolightmeditechusa.com

J/M9000vet-028-2011A2

Содержание M8000AVET

Страница 1: ...M8000VET Veterinary MonitorUser s Manual BIOLIGHT MEDITECH USA 535 Broadhollow Road Ste B 24 Melville New York 747 Http www biolightmeditechusa com J M9000vet 028 2011A2...

Страница 2: ...esult there might be cases of discrepancies between the manual and the actual situation of products When such discrepancies occur we will try our best to amend or add materials Your comments and sugge...

Страница 3: ...ument All the installation expansion readjustment renovation or repairs are conducted by the personnel certified by manufacturer The electrical safety status at the installation site of the instrument...

Страница 4: ...OR SAFETY 3 2 3 CLASSIFICATIONS 5 2 4 SAFE OPERATING AND HANDLING CONDITIONS 5 CHAPTER 3 PREPARATIONS BEFORE THE USE OF THE MONITOR 1 3 1 UNPACKING THE CASE 1 3 2 CONNECTING TO POWER 1 3 3 CONNECTING...

Страница 5: ...ALARM MODES 1 6 3 ALARM SETUP 2 6 4 ALARM CAUSE 4 6 5 SILENCE SUSPENSION 5 6 6 PARAMETER ALARM 5 6 7 WHEN AN ALARM OCCURS 6 6 8 ALARM DESCRIPTION AND PROMPT 6 CHAPTER 7 RECORDING 13 CHAPTER 8 THE MAI...

Страница 6: ...monitor specifications and performance criteria of the monitor the ambient requirements for the working and storage of the monitor Chapter 2 gives important safety notes Please do read this chapter b...

Страница 7: ...ust be strictly followed so as to prevent the operator or the veterinary patient from being harmed Caution Means it must be followed so as not to damage the instrument Note Important information or in...

Страница 8: ...pieces of alarm events Function of NIBP measurement reviewing can store 750 1000 pieces of NIBP measurement data Function of reviewing 10 30 minutes one important lead s EGC waveform Built in recorder...

Страница 9: ...000VET veterinary monitor Fig 1 4 2 The appearance of M9000VET veterinary monitor Various kinds of sockets See Fig 1 5 1 Recorder Alarm light Trim Knob See Fig 1 6 1 Function button zone See Fig 1 6 1...

Страница 10: ...ution The AC input socket at the back panel of the monitor can be connected with 100 240V AC power by electrical wires supplied with this instrument Various kinds of sockets See Fig 1 5 1 Recorder Tri...

Страница 11: ...RS 232 is only used for maintenance and upgrading of the monitor by manufacturer Note The Network Connector is a standard RJ45 socket and being used for connection with the central monitoring system p...

Страница 12: ...his symbol contains a F Type isolated floating applied part providing a high degree of protection against shock and is defibrillator proof Attention Please refer to the document supplied with this ins...

Страница 13: ...battery is used as the power source this indicating light is orange CHARGE Indicating light of CHARGE When the monitor is connected to the AC power of charge this indicating light is turn on When the...

Страница 14: ...audio system or cancel the silence NIBP STAT Switching type button Begin or Stop the measurement of NIBP Begin the STAT Press this button in 2 seconds to start or stop the NIBP measurement Press this...

Страница 15: ...Size M9000Vet 12 1 M8000Vet M8000AVET 10 4 Type Color TFT LCD Resolution 800 600 pixels or higher Indicators Alarm LED 1 Yellow Red AC Power LED 1 Green Orange Battery Charge LED 1 Yellow System outpu...

Страница 16: ...ure Operating temperature 0 40 Transportation and storage temperature 20 55 Relative humidity Working 85 Transportation and storage 93 Atmospheric pressure Working 860 1060 hPa Transportation and stor...

Страница 17: ...nd and light alarming Time to Alarm for Tachycardia Average 4s Tall T Wave Rejection Capability 0 1mV T Wave amplitude Response Time of Heart Rate Meter to Change in Heart Rate HR change from 80 to 12...

Страница 18: ...inuously adjustable between upper limit and lower limit MEAN 0 300 mmHg continuously adjustable between upper limit and lower limit Alarm indication Sound and light alarming Measurement Mode Horse Man...

Страница 19: ...Measurement Range 0 0 50 0 Accuracy 0 1 Resolution 0 1 Unit Celsius Fahrenheit Refreshing time 1s Self check Every 10 minutes Accuracy At 45 1 50 0 0 2 exclusive of probe At 25 0 45 0 0 1 exclusive o...

Страница 20: ...Static Dynamic 2mmHg or 2 of the reading whichever is greater exclusive of transducer 4mmHg or 4 of the reading whichever is greater inclusion of transducer 4mmHg or 4 of the reading whichever is grea...

Страница 21: ...on rate measurement 3 150 rpm Calibration Offset calibration auto manual Gain calibration Range of alarm 0 0 13 1 0 99 6mmHg continuously adjustable between upper limit and lower limit Alarm indicatio...

Страница 22: ...rature of 25 C 0 40 mmHg 2 mmHg 41 70 mmHg 5 of reading 71 100 mmHg 8 of reading 101 150 mmHg 10 of reading Above 80 breath per minute 12 of reading Gas temperature at 25 CO2 response time 3s includes...

Страница 23: ...O ENF SEV DES Resolution 1mmHg Unit mmHg Calibration Room air calibration performed automatically when changing airway adapter 5s Warm up time Concentrations reported in less than 10s full accuracy wi...

Страница 24: ...Sound and light indication Recorder Option Method Thermal dot array Paper width 50 mm Record width 40 mm Paper Speed 12 5 mm s 25 mm s 50 mm s Traces Maximum 3 tracks Alarm Level Low medium and high...

Страница 25: ...r can be used Connect the electrical wire to a properly grounded socket Avoid putting the socket used for it in the same loop of such devices as the air conditioners which regularly switch between ON...

Страница 26: ...trode of chest lead and move the electrode of limb lead to the side of the limb The electrode of the defibrillator should not come into direct contact with the monitoring electrodes Please ensure the...

Страница 27: ...rvice technician before it is used again ACCURACY If the accuracy of any value displayed on the monitor or printed on a printout paper is questionable determine the patient s vital signs by alternativ...

Страница 28: ...rs At the end of its service life the product described in this manual as well as its accessories must be disposed of in compliance with the guidelines regulating the disposal of such products If you...

Страница 29: ...Type CF applied part Not suitable Equipment is not suitable for use in the presence of flammable anesthetic mixture with air or with oxygen or nitrous oxide 2 4 Safe Operating and Handling Conditions...

Страница 30: ...the monitor and accessories Check the monitor and its accessories one by one in accordance with the particular paper Check to see if the parts have any mechanical damages In case of problems please co...

Страница 31: ...battery You must charge the battery before using it There is no external charger The battery is charged when the monitor is connected to AC power To assure a fully charged battery that is ready for us...

Страница 32: ...nt to the bottom of the monitor 4 After pushing the battery inside the storage withdraw the baffle turn back to the middle position 5 Close the gate Uninstall battery 1 Open the battery gate according...

Страница 33: ...as possible 3 5 Connecting to Various Kinds of Sensors Connect sensor cables to the relevant sockets on the monitor and put sensors on the monitored locations on the body of the patient Refer to the...

Страница 34: ...ng the steps to install recording paper 1 Push down the switch to open recorder 2 Install the paper with the thermal side upwards 3 Close the recorder with a section of paper outside of the storage Fo...

Страница 35: ...rs include 5 lead ECG RESP SpO2 Single TEMP and NIBP and the optional parameters include 2 channel TEMP 2 channel IBP EtCO2 and AG Note The monitor applies to large animals medium size animals and sma...

Страница 36: ...is set by the ECG1 option in ECG and the waveforms below are displayed differently according to different configurations 2 NIBP Review The recent groups of NIBP measurement results are displayed below...

Страница 37: ...Short Trend The short trend diagram relevant to the parameters is displayed on the upper left corner of the waveform 5 7 Leads The ECG waveforms of 7 lead are displayed in the waveform display zone th...

Страница 38: ...ne waveform and parts of parameters Among them through Bed NO the number of online machine can be selected and through Bed wave the waveform display of other beds can be selected Press Run to initiate...

Страница 39: ...Setup Click and open the dialog of alarm configuration Conduct configuration of alarm parameters New Patient Terminate the monitoring of the current patient and initiate the monitoring of a new patien...

Страница 40: ...onfiguration Some of the modules not in current use can be switched off and after switching off the relevant parameters and waveforms will not be displayed and no alarm will be made Trend storage Clic...

Страница 41: ...veform 1 Waveform 3 Select the waveform displayed in the third line Select Off close the wave display or select certain waveform to display Waveform 4 Select the waveform displayed in the fourth line...

Страница 42: ...ameters and relevant alarm will no be displayed and the current RESP waveform will be automatically switched off After it is open if there is no CO2 module the RESP waveform will be opened automatical...

Страница 43: ...e Click and open the dialog of IBP module setup IBP1 module Enable Disable the display of IBP1 module After switching off no IBP1 parameters and relevant alarm will be displayed and the current IBP1 w...

Страница 44: ...en it is enabled it indicates after NIBP measurement completed a record will be stored Alarm storage Enable Disable the switch of alarm storage When it is enabled it indicates if there is a high alarm...

Страница 45: ...mm IBP1 scale Select the scale of IBP1 for short trend diagram Options are 0 300mmHg 0 150mmHg 0 200mmHg 0 100mmHg 20 50mmHg and 50 300mmHg IBP2 scale Select the scale of IBP2 for short trend diagram...

Страница 46: ...in the third line Select Off close the wave display or select certain waveform to display Record Time Select the time duration of the waveform for each recording Options are 8s 12s and 16s Record int...

Страница 47: ...advised to switch off the monitor after exiting the current window and then restart it The time for the revision takes effect after the current window is exited Mode Setup Default Config Select the d...

Страница 48: ...stom configuration that needs to be deleted press the selected mode to delete the mode and press Cancel to cancel deleting Caution The mode name cannot be black when saving current configuration other...

Страница 49: ...2 Cal Mode Open or close the CO2 calibration mode When conducting calibration on sidestream CO2 set the CO2 cal mode to ON HUM Select the frequency of the AC power supply and options are 50Hz and 60Hz...

Страница 50: ...k Setup Click and open the dialog of network configuration Conduct network configurations Over press Initiate NIBP over pressure test Manometer Initiates NIBP manometer test NIBP reset Reset NIBP modu...

Страница 51: ...ams and on each page trend diagram can be configured for six regions and options are Off HR SpO2 NIBP PR Resp CO2 T1 T2 AA N2O O2 P1 P2 ST HR SpO2 SpO2 PR Resp CO2 PR CO2 T1 T2 P1 P2 AA CO2 N2O O2 It...

Страница 52: ...red for six regions and options are HR SpO2 NIBP S D NIBP M IBP1 S D IBP1 M IBP2 S D IBP2 M Resp PR T1 T2 CO2 AA N2O O2 ST It is possible to have self configurations on the contents of the trend table...

Страница 53: ...dress Net mask Gateway Machine number can be configured The configuration is mainly necessary when the monitor connecting to the Central Unit System info Version It displays the version number of soft...

Страница 54: ...ween the trend diagrams or trend tables between different pages Cursor Press this option turn the trim knob and move the cursor in the trend diagrams or trend tables Press it again to restore the init...

Страница 55: ...h 10h 12h 24h 48h and 72h Graph Press this option to switch to the display of trend diagram Table Press this option to switch to the display of trend tables 4 2 3 Alarm Review Select this button turn...

Страница 56: ...Chapter 4 22 4 2 4 Alarm Setup Common Alarm Click and open the dialog of common parameters alarm It can setup the alarm limits of common parameters IBPAlarm Click and open the dialog of IBP alarm It c...

Страница 57: ...ule Alarm Record Click and open the dialog of alarm recording Configure whether the alarm records of various modules are recorded Only when the switch for alarm recording of the module and the switch...

Страница 58: ...on press the button Adjust Alarm and the cursor moves to the adjustment region of alarm limits Press the button Enable All and all the alarms will be opened If the user desires to adjust the alarm par...

Страница 59: ...trim knob with an increment or decrement of 1 year Room No Number of patient s room Patient s room number can be displayed in the central unit Bed No Number of patient s bed Patient s bed number can b...

Страница 60: ...entration Turn the trim knob to move the cursor to the positions of the various calculation items in the calculation formula respectively turn the trim knob and select calculation value then press the...

Страница 61: ...e of titration table In the titration table turn the trim knob to Base then press the trim knob to select the desired item Options are Dose Trans speed and Drop speed After selecting press the trim kn...

Страница 62: ...tery capacity is half full Battery capacity is exhausted Only when the monitor is powered by battery and is recharging the battery the icon for battery capacity is displayed If AC power in current use...

Страница 63: ...omatically cancelled Pause the alarm and if a new alarm is generated the Pause status of alarm sound will be automatically cancelled Alarm indicating zone Alarm levels Red base color is high alarm Yel...

Страница 64: ...rative to only use the ECG cables provided with the instrument by manufacturer Warning The equipment is capable of displaying the ECG signal in the presence of pacemaker pulses without rejecting pacem...

Страница 65: ...rrent is limited to the safety range according to standards IEC 60601 2 27 5 1 3 Preparatory Steps before the Measurement of ECG HR 1 Plug the ECG cable into the ECG socket of the monitor 2 Place the...

Страница 66: ...lead attachment site with alcohol then place paste on the moistened hair and skin It is important that the paste be in direct contact with skin For patients with dense undercoat rub paste with fingers...

Страница 67: ...w Left foreleg RL Green N Black Right hind leg LL Red F Green Left hind leg V Brown C White 4th Intercostal Space left When conducting 3 leads ECG monitoring use 3 lead ECG cable The three limb leads...

Страница 68: ...Select the gain item of ECG waveform and options are AUTO 0 25x 0 5x 1 0x 2 0x and 4 0x HR source Select HR source item and options are AUTO ECG and PLETH Beep Volume Select the volume of BEEP and op...

Страница 69: ...limit and the lower part is the low limit The configuration range of high limit is 0 350 bpm continuously adjustable not lower than low limit and the configuration range of low limit is 0 350 bpm con...

Страница 70: ...ific frequencies 50HZ 60HZ are configured in Machine Setup and they must be configured according to the frequency of local power supply Display PR Select to simultaneity display pulse rate If simultan...

Страница 71: ...ad input mode and Lead I II III aVR aVL and aVF and one chest lead can be measured at the same time 5 1 7 Maintenance and Cleaning If there is any sign that the ECG cable may be damaged or deteriorate...

Страница 72: ...itoring 5 2 3 Connect the ECG Cable with Patient and the Monitor To measure RESP parameters it is unnecessary to use other cables and it is only necessary to use the two RA and LL leads in the 5 lead...

Страница 73: ...When the system is configured with CO2 module RESP source can be selected as AUTO ECG and EtCO2 Only when the monitor that user has bought has CO2 module EtCO2 of RESP source is valid otherwise the R...

Страница 74: ...after the light passes the capillary network and estimates the ratio of hemoferrum and the total hemoglobin Degree of pulse oxygen saturation globin desoxyhemo hemoferrum hemoferrum 100 Abnormal hemog...

Страница 75: ...Under such conditions as shock hypothermia anemia or the use of blood vessel activating drugs and with the existence of such substances as carboxyhemoglobin methemoglobin methylene blue the result of...

Страница 76: ...utomatically derives HR from SpO2 Scan speed Select the scanning speed of the ECG waveform and options are 12 5mm s 25mm s and 50mm s Alarm Setup Click and open the dialog of SpO2 alarm configuration...

Страница 77: ...Note When disposing the disposable SpO2 probe or useless SpO2 probe please observe all local state and federal regulations that relate to the disposal of this products or similar products For reusable...

Страница 78: ...he TEMP sensors on body of patient and the screen will show the value of TEMP measurement 3 Set the parameters relevant to TEMP 5 4 3 Connecting Patient and Monitor Refer to Fig 1 5 1 and plug the TEM...

Страница 79: ...figuration range is 0 50 continuously adjustable no lower than the low limit the configuration range of TEMP2 alarm low limit is 0 50 continuously adjustable no higher than the high limit T1 alarm Sel...

Страница 80: ...f test of the temperature measurement is performed automatically once every 10 minutes during the monitoring The test procedure lasts about one second and does not affect the normal measurement of the...

Страница 81: ...counterclockwise first before unplugging 2 Place the cuff on the veterinary patient Place the patient on a padded surface or chair to provide comfort Shivering will inhibit the monitor from making a d...

Страница 82: ...so the metacarpal area of the forelimb is most convenient In situations where this is not possible the cuff should be wrapped around the metatarsus just proximal to the tarsal pad or around the hind l...

Страница 83: ...osition of the cuff is higher than the plane of heart the measured BP readings tend to be smaller in case the position of the cuff is lower than the plane of the heart the measured BP readings tend to...

Страница 84: ...s confirmed Object Objects of measurements shall be configured and options are horse dog and cat When monitoring a cat or small animal set the object to cat when monitoring dogs or medium size animals...

Страница 85: ...igh limit Mean alarm high limit its configuration range is 0 300mmHg continuously adjustable no lower than the low limit the configuration range of Mean alarm low limit is 0 300mmHg continuously adjus...

Страница 86: ...ring individual blood pressure readings taken over time Normal feline blood pressure 124 84 The normal values for dogs are breed specific Those for Golden Retrievers Labradors and giant breeds tend to...

Страница 87: ...rn the trim knob to the Manometer option and press Then the NIBP module has started performing calibration 5 Inflate the pneumatic system to 0 50 and 200 mmHg by ball pump separately The difference be...

Страница 88: ...the manufacturer for repair 8 Press the NIBP STAT button on front panel can also stop the test Fig 5 5 4 Diagram of air leakage check 5 5 8 Maintenance and Cleaning Warning Do not squeeze the rubber...

Страница 89: ...of the continuous BP The IBP parameters of Monitor can select Arterial Pressure ART Pulmonary Artery pressure PA Left Atrium Pressure LAP Right Atrium Pressure RAP Central Venous Pressure CVP Intracr...

Страница 90: ...r the defibrillation the monitoring will go on normally the operation mode and the user configuration are not affected Warning The operator should avoid contact with the conductive parts of the appurt...

Страница 91: ...er one is the low limit Sys alarm high limit its configuration range is 50 300mmHg continuously adjustable no lower than the low limit the configuration range of Sys alarm low limit is 50 300mmHg cont...

Страница 92: ...In the course of zeroing should turn off the three way valve near artery needle don t connect artery needle with patient and make sure there is no air inside the whole tube 5 6 5 Connecting to Patient...

Страница 93: ...Transducer Zero and Calibration IBP Transducer Zero Warning It is the responsibility of the user to ensure that a zero procedure has recently been done on the transducer otherwise there will be no re...

Страница 94: ...at would lead to patient is off 4 Connect the 3 way connector to the 3 way stopcock that is not connected to the patient catheter 5 Open the port of the 3 way stopcock to the sphygmomanometer 6 Inflat...

Страница 95: ...obe depends on the concentration of the measured CO2 5 7 1 Brief Introduction to Measurement of Sidestream EtCO2 According to the Fig 5 7 1 snap slantways dehydration flask on the receptacle fixed on...

Страница 96: ...etup button in Main Screen then select the Start and press this button to start sampling pump begin measuring EtCO2 Pay attention to the water level of dehydration flask If the highest water level rea...

Страница 97: ...O2 alarm Start Press this button to start the sampling pump to initiate the measurement of CO2 only valid on sidestream CO2 Stop Press this button to switch off the sampling pump and terminate the mea...

Страница 98: ...lower than the low limit the configuration range of FiCO2 alarm low limit is 0 0 13 1 0 99 6mmHg continuously adjustable no higher than the high limit RESP alarm high limit its configuration range is...

Страница 99: ...ied out if the previous conditions happened The following procedures must be observed 1 In parameter setup please adjust VIEW TYPE of the CO2 module to InsCO2 When the monitor has been run for 30 minu...

Страница 100: ...ading 5 8 Measurement of CO2 Mainstream IRMA Note You can only use PHASEIN IRMA mainstream EtCO2 probe provided by the manufacturer to perform EtCO2 monitoring on the monitor 5 8 1 Preparatory Steps f...

Страница 101: ...he oxygen sensor will be performed automatically at regular intervals whenever the IRMA sensor head is disconnected from the IRMA airway adapter If the IRMA sensor is kept in operation for a long time...

Страница 102: ...to Main Return to main screen 5 8 6 Precautions during Measurement 1 Plug IRMA probe into the CO2 socket on the left side panel of the monitor then connect IRMA airway adapter to the breathing circui...

Страница 103: ...lar intervals A zero reference calibration of the IR measurement should be performed whenever an offset in gas readings is discovered or if GAS CONC OUT OF RANGE alarms appear when measuring room air...

Страница 104: ...cable plug the cable into the CO2 socket on the left side panel of the monitor by matching the key on the cable to the key on the connector Fig 5 9 1 Fig 5 9 1 Caution To remove the module cable from...

Страница 105: ...nsor Warm Up message for approximately one minute while the module warms up to operating temperature The message disappears when the module is ready for use Note Warm up time varies with ambient tempe...

Страница 106: ...sly adjustable no lower than the low limit the configuration range of FiCO2 alarm low limit is 0 0 13 1 0 99 6mmHg continuously adjustable no higher than the high limit RESP alarm high limit its confi...

Страница 107: ...gen Turn the trim knob with an increment or decrement of 1 Balance gas Select the balance gas type the options are Air N20 and Helium Anesthetic Select the concentration of balance gas Turn the trim k...

Страница 108: ...r or cannula from the patient s airway This time allows any CO2 remaining in the adapter or cannula to dissipate before zeroing Do not attempt to zero the module while the adapter or cannula is in the...

Страница 109: ...r or sample line from the sensor when not in use 5 9 5 Removing Exhaust Gases from the System Warning When using the microstream CO2 measurement on patients who are receiving or have recently received...

Страница 110: ...airway adapters and the sampling line Replace it if needed Do not operate the LoFlo Module when it is wet or has exterior condensation Do not apply excessive tension to any cable Do not use device on...

Страница 111: ...ed on the patient s size ET tube diameter and monitoring situation 3 Attaching the airway adapter to the CAPNOSTAT 5 sensor Before attaching the airway adapter to the CAPNOSTAT 5 sensor verify that th...

Страница 112: ...damaged Refer servicing to qualified service personnel Do not position the sensor cables or tubing in any manner that may cause entanglement or strangulation Reuse disassembly cleaning disinfecting or...

Страница 113: ...OSTAT5 sensor and its accessories have been met disposal should be accomplished following national and local requirements Nitrous oxide elevated levels of oxygen and helium can influence the CO2 measu...

Страница 114: ...optic infrared filter selects the infrared light with special wavelength to penetrate this gas For a given volume the higher is the gas concentration the more infrared light is absorbed We may measur...

Страница 115: ...monitor display 5 11 3 Room Air Calibration Room air calibration of the oxygen sensor will be performed automatically at regular intervals whenever the IRMA sensor head is disconnected from the IRMA...

Страница 116: ...ypes of anesthetic gas and options are AA HAL ENF ISO SEV and DES After the monitor is turn on if no AA types are configured there will be a technical alarm prompting the configuration of AA and need...

Страница 117: ...the low limit the configuration range of EtAA alarm low limit is 0 0 30 0 continuously adjustable no higher than the high limit FiAA alarm high limit its configuration range is 0 0 30 0 continuously...

Страница 118: ...is 18 100 continuously adjustable no higher than the high limit FiO2 alarm cannot be switched off and when lower than 18 it will trigger high alarm EtO2 alarm high limit its configuration range is 10...

Страница 119: ...the low limit the configuration range of EtN2O alarm low limit is 0 100 continuously adjustable no higher than the high limit FiN2O alarm Select ON to enable FiN2O over limit alarm select OFF to disab...

Страница 120: ...up for at least 15 minutes after power on and 2 minutes after changing airway adapter before transmitting the calibration command Zero Reference calibration is performed by snapping a new IRMA airway...

Страница 121: ...iority and low priority High priority alarm indicates the patient s life is in danger It is the most serious alarm Medium priority alarm means serious warning Low priority alarm is a general warning O...

Страница 122: ...ndicating lamp flashes in yellow with 0 5 Hz Low Alarm indicating lamp lights on in yellow Screen Display Physiological alarm The parameter which triggers the alarm splashes in the frequency of 2Hz on...

Страница 123: ...I Please refer to above operation for Methods of Alarm setup of the other parameters It is important to set physiological alarm limits properly The monitor can t give medicinal alarm prompt in clinica...

Страница 124: ...sed on clinical practices and general clinical experiences Note When parameter alarm level is off alarm will be disabled even if the measurement results exceed the limits Alarm indicating lamp in the...

Страница 125: ...ke the system back to normal status When in the SILENCE status the icon will be displayed in the left upper of the screen SUSPENSION Press the SUSPEND SILENCE button on the front panel for less than 2...

Страница 126: ...recorder starts alarm recording at set interval 6 7 When an Alarm Occurs Note When an alarm occurs you should always check the patient s condition first Check the alarm message appeared on the screen...

Страница 127: ...th in specific interval User Selectable Technical Alarm Message Cause Alarm Level RR alarm error Alarm failure Low 6 8 3 SpO2 Alarm Physiological Alarm Message Cause Alarm Level SpO2 too high SpO2 mea...

Страница 128: ...larm limit User Selectable Technical Alarm Message Cause Alarm Level TEMP1 sensor off TEMP1 sensor may be disconnected from user or monitor Low TEMP2 sensor off TEMP2 sensor may be disconnected from u...

Страница 129: ...ring value of channel 2 is below lower alarm limit User Selectable IBP MAP2 too high MAP measuring value of channel 2 is above upper alarm limit User Selectable IBP MAP2 too low MAP measuring value of...

Страница 130: ...surement failure or communication failure Low NIBP SYS alarm error Alarm failure Low NIBP DIA alarm error Alarm failure Low NIBP MAP alarm error Alarm failure Low Technical Alarm 2 display in descript...

Страница 131: ...e Cause Alarm Level NIBP Resetting NIBP measurement module is resetting No alarm Over Press Testing NIBP is testing Over Pressure Manometer Testing NIBP is testing Manometer Pneumatic Testing NIBP is...

Страница 132: ...Lo EtCO2 measuring value is below lower alarm User Selectable FiCO2 Hi FiCO2 measuring value is above upper alarm User Selectable FiCO2 Lo FiCO2 measuring value is below lower alarm User Selectable Ap...

Страница 133: ...e FiN2O too low FiN2O is below lower alarm limit EtO2 too high EtO2 is above upper alarm limit User Selectable EtO2 too low EtO2 is below lower alarm limit FiO2 too high FiO2 is above upper alarm limi...

Страница 134: ...er can adjust Record Interval of Recorder Setup of System Setup in Monitor Setup to a necessary interval time All monitoring parameter values and waveforms will be recorded automatically according to...

Страница 135: ...o patient monitoring make sure that the monitor is in good condition If you find any damage on the monitor stop using the monitor on patient and contact the biomedical engineer of the hospital or Manu...

Страница 136: ...ot follow a satisfactory maintenance schedule the monitor may become invalid and the human health may be endangered Note To ensure maximum battery life please ensure that the battery is fully charged...

Страница 137: ...on battery in need of conditioning into battery compartment of the monitor 3 Connect the monitor to the AC mains Allow the battery to be charged uninterruptedly for above 6 hours 4 Remove the AC mains...

Страница 138: ...to be cleaned Hydrogen Peroxide 3 Alcohol 70 Isopropanol 70 The surface of patient monitor can be cleaned with hospital grade ethanol and dried in air or with crisp and clean cloth The manufacturer h...

Страница 139: ...rture ECG CABLE 5 lead ECG cable 6pin snap IEC BLT EABF 3 lead ECG cable 6pin snap IEC BLT EAAF 5 lead ECG cable 6pin snap AHA BLT EABA 3 lead ECG cable 6pin snap AHA BLT EAAA 5 lead ECG cable 6pin 4m...

Страница 140: ...tion Temperature Probe 90044 YSI 400 Series 5 IBP Accessory Description IBP Transducer Deltran II DPT 248 6 CO2 Accessory Description EtCO2 Sidestream CPT Water trap Sample line 3 way stopcock EtCO2 M...

Страница 141: ...Veterinary monitor user s manual Chapter 9 3 AG IRMA IRMA AX sensor Extension cable Airway adapter...

Страница 142: ...Resp Region 6 T1 T2 2 Trend Table Configuration Page 1 Region Parameter Region 1 HR Region 2 SpO2 Region 3 PR Region 4 NIBP S D Region 5 NIBP M Region 6 Resp Page 2 Region Parameter Region 1 HR Region...

Страница 143: ...n dual IBP EtCO2 1 Trend Graph Configuration Page 1 Region Parameter Region 1 HR Region 2 SpO2 Region 3 P1 Region 4 P2 Region 5 CO2 Page 2 Region Parameter Region 1 PR Region 2 NIBP Region 3 Resp Regi...

Страница 144: ...egion Parameter Region 1 PR Region 2 NIBP Region 3 Resp Region 4 O2 N2O Region 5 AA Region 6 T1 T2 2 Trend Table Configuration Page 1 Region Parameter Region 1 HR Region 2 SpO2 Region 3 P1 S D Region...

Страница 145: ...IA mmHg 20 70 NIBP MEAN mmHg 60 90 IBP1 SYS mmHg 80 130 IBP1 DIA mmHg 20 70 IBP1 MEAN mmHg 60 90 IBP2 SYS mmHg 80 130 IBP2 DIA mmHg 20 70 IBP2 MEAN mmHg 60 90 EtCO2 2 7 8 1 FiCO2 0 1 1 EtAA 0 0 3 0 Fi...

Страница 146: ...FiAA 0 0 5 0 EtN20 0 82 FiN20 0 82 EtO2 10 100 FiO2 18 100 3 Setup of parameters alarm limit for CAT Parameter Low limit High limit HR bpm 90 200 SpO2 90 100 PR bpm 90 200 RR rpm 8 40 T1 38 1 39 2 T2...

Страница 147: ...Veterinary monitor user s manual 6 FiAA 0 0 5 0 EtN20 0 82 FiN20 0 82 EtO2 10 100 FiO2 18 100...

Страница 148: ...he customer or the user of Low Frequency Therapeutic Device should assure that it is used in such an environment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidanc...

Страница 149: ...separation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interfe...

Страница 150: ...ecommended below according to the maximum output power of the communications equipment Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz t...

Страница 151: ...y Monitor Product type M8000VET Address 535 Broadhollow Road Ste B 24 Melville New York 747 Http www biolightmeditechusa com Tel 888 82 BLUSA 888 822 5887 Fax 888 785 8948 Email sales biolightmeditech...