background image

 

20 

 

Recommended separation distances between   

portable and mobile RF communications equipmen t and the device

 

The  device  is  intended  for  use  in  an  electromagnetic  environment  in  which  radiated  RF  disturbances 

are controlled. The customer or the user of the  device can help prevent electromagnetic interference by 

maintaining  a  minimum  distance  between  portable  and  mobile  RF  communications  equipment 

(transmitters)  and  the  device  as  recommended  below,  according  to  the  maximum  output  power  of  the 

communications equipment. 

Rated maximum 

output power of 

transmitter 

(W) 

Separation distance according to frequency of transmitter 

(m) 

150 KHz to 80 MHz 

80 MHz to 800 MHz 

800 MHz to 2.5 GHz 

0.01 

0.12 

0.12 

0.23 

0.1 

0.38 

0.38 

0.73 

1.2 

1.2 

2.3 

10 

3.8 

3.8 

7.3 

100 

12 

12 

23 

For  transmitters  rated  at  a  maximum  output  power  not  listed  above,  the  recommended   separation 

distance  d  in  meters  (m)  can  be  estimated  using  the  equation  applicable  to  the  frequency  of  the 

transmitter,  where P is the  maximum output power rating of the  transmitter in watts (W) according to 

the transmitter manufacturer. 

 

NOTE 1      At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.  

NOTE  2      These  guidelines  may  not  apply  in  all  situations.  Electromagnetic  propagation  is  affected 

by absorption and reflection from structures, objects and people.  

 

Содержание WBP100

Страница 1: ...Electronic Sphygmomanometer User s Manual Guangdong Biolight Meditech Co Ltd ...

Страница 2: ...6 400 8818 233 Service life 5 years Revision History This manual has a revision number This revision number changes whenever the manual is updated due to software or technical specification change Contents of this manual are subject to change without prior notice Document No J WBPCE A 008 Revision number V1 0 Release time Sept 2015 Copyright 2015 Guangdong Biolight Meditech Co Ltd All rights reser...

Страница 3: ...lity and performance of the instrument All the installation expansion readjustment renovation or repairs are conducted by the personnel certified by manufacturer The storage condition operation condition and electrical status of the instrument conform to the product specification The instrument is used in accordance with the user s manual About this manual This manual contains the instructions nec...

Страница 4: ...l hazard or unsafe practice that if not avoided will result in death or serious injury Caution Indicates a potential hazard or unsafe practice that if not avoided could result in minor personal injury or product property damage Note Provides application tips or other useful information to ensure that you get the most from your product ...

Страница 5: ...5 Display 6 6 Setting 7 7 Historical data review function 8 8 Download And Install The Software 8 9 Using Steps 8 10 Maintenance and Cleaning 11 11 Accessory and Specifications 12 12 Troubleshooting 14 13 Equipment Symbols 15 14 Blood pressure general knowledge 15 Appendix A Guidance and Manufacturer s Declaration of EMC 17 ...

Страница 6: ...e is reusable and intended for measuring human blood pressure and pulse rate Application population Adult Measurement site Place a bare arm through the cuff and position the cuff 1 2cm above the elbow joint 3 Safety Information Warning Don t take measurement immediately after smoking drinking alcohol coffee and black tea or within an hour after a meal Magnetic and electrical fields are capable of ...

Страница 7: ... This product might not meet its performance specifications if stored or used outside the specified temperature humidity and altitude ranges Please do not share the cuff with any infectious person to avoid cross infection Battery replacement should only be performed by personnel who have been confirmed qualified by the manufacturer To do otherwise will void your warranty and possible damage your u...

Страница 8: ... the user s authority to operate the equipment This equipment has been tested and found to comply with the limits for a Class B digital device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference in a residential installation This equipment generates uses and can radiate radio frequency energy and if not installed and used in ...

Страница 9: ...d range specified in part specifications the device will immediately display an error code on screen In this case consult a physician or ensure that proper measurement procedures are followed Patients can use all functions of the device safely according to this user s manual Any questions about setting using or maintaining the device please contact with manufacturer If unexpected operation or even...

Страница 10: ...on 3 Power button Start Stop button Long press press and hold it more than 2s this button to turn off the device When the power is off short press press and hold it less than 2s this button to turn on the device Otherwise short press it to start stop the NIBP measurement 4 Charging cable interface Connect to the power supply for charging 1 3 4 5 5 6 4 3 2 1 ...

Страница 11: ...nit 3 Diastolic blood pressure 4 Pulse rate 5 Time 6 Users name 7 Pulse rate unit 8 Signal quality 9 Bluetooth state 10 Battery state When measuring error such as transducer failure loosely wrapped cuff screen displays the wrong code The definition of the error code and the possible reason are as follows Error code Definition Possible Reason E02 Self test failed Transducer or other hardware failur...

Страница 12: ...torical data review button and hold it for 3 seconds to enter the year setting interface The third number of the year starts to flash When you press the select users button once the third number increase 1 It means the year increases ten years After you finish setting the third number of year press the historical data review button to enter the setting interface of the fourth number The fourth num...

Страница 13: ... the searching results Users who use the Android system device can log in the wearcare official website http www wearcare cn for download 2 Download the software of BP tracker click installation button for free installation 9 Using Steps Preparation before measurement 1 Keep your body and mood calm take off the thicker clothes such as coat sweater Etc When measuring bare upper arm or wear thinner ...

Страница 14: ...ably yet snugly around your arm You should be able to insert one finger between your arm and the cuff Caution Make sure that the appropriate cuff size is used Refer to the cuff circumference range in the Specification section of this manual 2 Body posture Stay still and calm for five minutes before taking a blood pressure measurement Be seated with your feet flat on the floor without crossing your...

Страница 15: ... use the APP software can read the measurement results on the software interface 5 Take off the cuff 6 Turn off the device Long press the power button to turn off the device after the measurement Note While measuring blood pressure the users must stay still and calm without any talk The cuff tied on the limb shall be on the same level as the patient s heart so as to avoid the reading error resulti...

Страница 16: ...s must be diluted before use Please use them according to the directions of the manufacturers of the detergents 3 The casing of the device should be kept from the contamination of filth and dirt and it can be wiped with non velvet soft cloth When cleaning do not spill the liquid onto the instrument Ensure no liquid is allowed to enter the inside of the device 4 It is recommended that if the cuff i...

Страница 17: ...nt list are only provided to repair station or personnel which have been confirmed qualified by the manufacturer Check the device regularly please If you doubt the measurement result is not accurate please contact the manufacturer 11 Accessory and Specifications 11 1 Optional item Items Mode Power adapter optional LXCP12 05 Output DC5V 1 0A 11 2 Technical Specifications Production model WBP100 WBP...

Страница 18: ...d WBP 201 WBP 202 Bluetooth state indicating Not connected The indicating flashes Connected The indicating lights without flashing Battery state indicating The electric quantity of battery is divided into four ranks to display the current electricity Being Charged The indicating flashes Not being charged or fully charged The indicating lights without flashing Bluetooth Bluetooth version Bluetooth ...

Страница 19: ... no electricity 1 The device itself might be damaged 1 Please charge the device 2 Contact customer service The data cannot be display properly or display reads an abnormal result 1 The cuff position was not correct or it was not properly tightened 2 Speaking moving arm or body being angry excited or nervous during test 1 Review the cuff application instructions and retest 2 Retest when calm avoid ...

Страница 20: ...transport package contents fragile so handling should be handled with care Upward It shows the correct position of the transport package is upright Guard against wet Show packages afraid be wet Stacking layer limit Same packing maximum stacking layers N represents the number of layers limit N is 6 14 Blood pressure general knowledge 14 1 Definition of blood pressure Blood pressure is the pressure ...

Страница 21: ...essure DIA kPa mmHg Corresponding measure Hypotension 13 3 100 8 0 60 Ask physician for measuring Normal blood pressure 13 3 100 18 7 140 8 0 60 12 0 90 Self measure Mild Hypertension 18 7 140 21 3 160 12 0 90 13 3 100 Consult physician Moderate Hypertension 21 3 160 24 0 180 13 3 100 14 7 110 Consult physician Severe Hypertension 24 0 180 14 7 110 Danger Please see the doctor as soon as possible ...

Страница 22: ...l operation in the configuration in which it will be used Guidance and manufacture s declaration electromagnetic emission The device is intended for use in the electromagnetic environment specified below The customer of the user of the device should assure that it is used in such an environment Emission test Compliance Electromagnetic environment guidance RF emissions CISPR 11 Group 1 The device u...

Страница 23: ...rential mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 sec 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in ...

Страница 24: ...mpliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol Radiated RF IEC 61000 4 3 3 V m 80 MHz to 2 5 GHz 3 V m NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and p...

Страница 25: ...Separation distance according to frequency of transmitter m 150 KHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where...

Отзывы: