Electronic Sphygmomanometer User’s manual
6
5
、
Select users button
—
In measurement sate: Short press this button to select users;
—
In setting state: Short press this button, the time value would increase one.
6
、
Historical data review button
—
Not in setting state: Short press this button to review historical data. Long
press it to enter setting mode.
—
In setting state: Short press this button to switch the setting items. Long
press it to save the setting and quit setting mode.
5
、
Display
1.
Systolic blood pressure;
2.
Blood pressure unit;
3.
Diastolic blood pressure;
4.
Pulse rate;
5.
Time;
6.
Users name;
7.
Pulse rate unit;
8.
Signal quality;
9.
Bluetooth state;
10.
Battery state.
When measuring error, such as transducer failure, loosely wrapped cuff, screen
displays the wrong code. The definition of the error code and the possible reason are
as follows:
Error code
Definition
Possible Reason
E02
Self-test failed
Transducer or other hardware failure.
E06
Loose cuff
a.
Cuff is completely unwrapped;
b.
No cuff attached;
E07
Air leak
Air leak in pneumatics, hose, or cuff.
E08
Air pressure error
Unit cannot maintain stable cuff pressure, e.g. kinked.
E09
Weak signal
a.
Very weak patient signal due to a loosely wrapped
cuff;
b.
Extremely weak pulse from patient.