Electronic Sphygmomanometer User’s manual
10
Note:
Incorrect measuring posture showed as follows might cause an incorrect
measurement:
a)
Bent over
(
lean forward
)
;
b)
Sitting cross-legged;
c)
Bend the body because of the low table or sitting on sofa.
(3)
Measuring
Select users: Press the Select users button to select users, you can select user
1 or user 2;
Start measuring: Press the power button, then the cuff pressure automatically
and the screen shows rising pressure values. Stay still and calm during
measurement. Do not move your arm, body and the device.
(4)
Display the measurement value
The device exhausts automatically and displays the measurement results. About
the display interface of measurement results, please refer to chapter “
5 Display
”.
Users, who use the APP software, can read the measurement results on the
software interface.
(5)
Take off the cuff
(6)
Turn off the device
Long press the power button to turn off the device after the measurement.
Note:
While measuring blood pressure, the users must stay still and calm without
any talk;
The cuff tied on the limb shall be on the same level as the patient’s heart so
as to avoid the reading error resulting from the hydrostatics effect of the blood
flow between the heart and cuff. If the cuff position is higher than heart level,
the BP reading will be lower, the measured result shall be added 0.75mmHg
(0.1kPa) for each centimeter higher; in case the cuff position is lower than
heart level, the BP reading will be higher, the measured result shall be deducted
0.75mmHg (0.1kPa) for each centimeter lower;
If the signal measurement time is more than two minutes and the device
doesn’t exhaust automatically, please take off the cuff and turn off the device.