Electronic Sphygmomanometer User’s manual
7
Error code
Definition
Possible Reason
E10
Range exceeded
Measurement range exceeds module specification of 255
mmHg maximum.
E11
Excessive motion
a.
Too many retries due to interference of motion
artifact;
b.
Signal is too noisy during step-down to detect
pressure pulse, e.g. subject has severe tremor;
c.
Irregular pulse rate, e.g. arrhythmia.
E12
Overpressure sensed
Cuff pressure exceeds software safety limit, 295 mmHg
in Adult mode. Could be due to rapid squeezing or
bumping of cuff.
E13
Signal saturated
Large motion artifact that saturates the BP amplifier’s
amplitude handling capability.
E14
Air system leak
Module report Air system leakage Failure while in the
Air System Leakage Test mode.
E15
System failure
Abnormal events.
E19
Time out
Measurement took more than 120 seconds in adult.
6
、
Setting
You can set the time and date of this product. The setting method is as follows:
Press the historical data review button “
”
and hold it for 3 seconds to enter
the year setting interface. The third number of the year starts to flash. When you press
the select users button “
”
once, the third number increase 1 (It means the year
increases ten years).
After you finish setting the third number of year, press the historical data review
button “
”
to enter the setting interface of the fourth number. The fourth number
of the year starts to flash. When you press the select users button “
”
once, the
fourth number increase 1 (It means the year increases 1 year).
Set month, day, hour, and minute in the same way.
After you finish setting time and date, press the historical data review button
“
”
to enter the setting interface of blood pressure units (The default unit is mmHg).
When a flashing mmHg appears on the screen, it means the blood pressure unit is
mmHg. Press the select users button “ ”, mmHg would disappear and kPa would
appear. Press the button “
”
again, kPa would disappear and mmHg would appear.
So repeatedly, you can switch between the two units. Press the historical data review
button “
”
and hold it for 3 seconds to end the setting after you finish setting unit.