Fig. 5

Fig. 4

5.4. Filters

Removing the nose cone exposes a hydrophobic 

fi

 lter (0.45 micron), located 

at the top of the nose piece. The 

fi

 lter protects the Midi Plus from accidental 

overaspirations. It is strongly recommended to use only the original 

fi

 lters 

supplied by the manufacturer with the Midi Plus. Both sterile and non-sterile 

fi

 lters are available (see 7.2. Changing the nose cone, silicon adapter or 

fi

 lter).

Note:

 The Midi Plus cannot be operated without a 

fi

 lter. Wetting of 

fi

 lter 

due to over

fi

 lling will slow or prevent 

fl

 uid movement. If over

fi

 lling occurs, 

change the 

fi

 lter immediately.

5.5. Support

The  Midi  Plus  also  features  an  attached 

support for placing the unit on the table 

between the dispensing routines with a 

pipet attached (Fig. 4).

6.  OPERATING THE MIDI PLUS

6.1. Adjusting the operating speed

Select the desired speed by turning the linear 

speed adjustment wheel (Fig. 5). The speed 

selection can be easily accessed as the height of 

the indicator column corresponds to the selected 

speed.

Note:

 The aspiration and dispensing speeds 

can be 

fi

 ne-tuned during operation by applying 

a varying amount of finger pressure to the 

operating buttons.

6.2. Inserting a pipet

Gently push a pipet into the silicone adapter and ensure that it 

fi

 ts securely. 

To insert the pipet safely, hold the Midi Plus from above the nose cone and 

the pipet from the top end of the tube.

Содержание Midi Plus

Страница 1: ...Midi Plus Instruction Manual Bedienungsanleitung Mode d emploi Manual Usaria Instruzioni d impiego Bruksanvisning K ytt ohje...

Страница 2: ...Instruction Manual 1 Bedienungsanleitung 13 Mode d emploi 25 Manual Usaria 37 Istruzioni d impiego 49 Bruksanvisning 61 K ytt ohje 73 Spare Parts See inside back cover BIOHIT Midi Plus...

Страница 3: ...ttons 5 5 2 Speed adjustment wheel 5 5 3 Pipet adapters 5 5 4 Filters 6 5 5 Support 6 6 OPERATING THE MIDI PLUS 6 6 1 Adjusting the operating speed 6 6 2 Inserting a pipet 6 6 3 Aspiration 7 6 4 Dispe...

Страница 4: ...ts The aspiration and dispensing speeds can also be fine tuned during operation by applying varying finger pressure to the operating buttons The Midi Plus has been designed so that it can be suspended...

Страница 5: ...transformer if it will be out of use for a lengthy period of time 3 2 Low battery indicator When the battery is close to empty the battery capacity indicator starts to blink Fig 2 The unit can still b...

Страница 6: ...SA PC Handle cover release buttons ASA PC Support PA Suspenison hook ASA PC 3 3 Electrical specifications Batteries AA 3 x 1 2 V 350 mAh NiCd Approx 7 hours service life when fully charged Charging ti...

Страница 7: ...ers The pipet adapter provided with the Midi Plus allows the use of all standard plastic and glass pipets graduated and non graduated including Pasteur pipets The pipet adapter of the Midi Plus consis...

Страница 8: ...mediately 5 5 Support The Midi Plus also features an attached support for placing the unit on the table between the dispensing routines with a pipet attached Fig 4 6 OPERATING THE MIDI PLUS 6 1 Adjust...

Страница 9: ...ut 2 Pressing the dispensing button further down engages the motor and activates the blow out function Note Finger pressure can be varied to fine tune the speed within each mode 6 5 Other pipetting re...

Страница 10: ...your Midi Plus 8 1 Decontamination To decontaminate lightly spray the outside of the Midi Plus with the Biohit Proline Biocontrol decontamination spray cat no 724004 5 l and wipe dry with a clean clo...

Страница 11: ...uid movement If overfilling occurs change the filter immediately Both sterile and non sterile filters are available Cat No Product Qty 712912 Filter for Biohit Midi Plus non sterile 5 pcs 712913 Filte...

Страница 12: ...VOID IF FAULT IS FOUND TO HAVE BEEN CAUSED BY MALTREATMENT MISUSE UNAUTHORIZED MAINTENANCE OF SERVICE OR NEGLIGENCE OF REGULAR MAINTENANCE AND SERVICE ACCIDENTAL DAMAGE INCORRECT STORAGE OR USE OF THE...

Страница 13: ...Directive 98 79 EC 11 RECYCLING INSTRUCTION WEEE In compliance with European Directive WEEE 2002 96EC on waste and reducting of hazardous substances in electical and electronic equipments this device...

Страница 14: ...12 Notes...

Страница 15: ...t 3 pcs 712991 Silicon adapter Standard 712893 10 ml 712911 5 ml 712910 Pump speed adjustment assembly 712111 Tubing set 712112 Pump assembly 712113 Connection valve 712114 Hydrophobic fi lter 0 45 m...

Страница 16: ...tions contenues dans cette publication sont jour au moment de la publication et ont t compil es avec le plus grand soin Biohit n est pas responsable des ventuelleserreursouomissionsetser serveledroitd...

Отзывы: