Gene Explorer Thermal Cyeler Operation Manual
Appendix 2
-
24
-
Appendix 2 Value and Meaning of Set Parameters
No.
Parameter
Meaning
Set value range
Remark
1
Temp(C)
Temperature control
point (C)
4.0~99.0
2
Time
Temperature control
time (mm:ss)
--:--
(press “+/-“ key before “:” to
set)
Infinite temperature control
time and turn off hot lid in
the meantime
00:00~59:59
3
Ramp(C/S)
Temperature heating
and cooling rate
(C/s)
0.1 ~4.0
#.#
Max. ramp rate
4
+Temp
Temperature
increase for each
cycle
(C)
0.0~9.9
5
+Time
Time increase for
each cycle
(m:ss)
0:00~9:59
6
Cycle1
Number of the cycle
1~5
Cycle1:×00 From 00 to 00
×00
Times of the cycle
1~99
From 00
Starting section of
the cycle
X(any existing section
number)
to 00
Ending section of
the cycle
≤X+15
7
Grad.
Gradient
temperature (C)
0 ~30
8
Date
Date (YY-MM-DD)
00(~99)-01(~12)-01(~31)
YY-MM-DD
9
Time
Time (HH:MM:SS)
00:00:00~23:59:59
HH:MM:SS
10
Default File
System default file
name
Max. 11 characters
11
Default User
System default user
name
Max. 11 characters
12
Control Mode
Temperature control
mode
Block
In this mode (enter into
constant temperature status
after
reaching
set
temperature), the time for
reagent at set temperature is
less than the set time (see
Fig. 1).
Tube
In this mode (enter into
constant temperature status
when overshoot for about
3
℃
after
reaching
set
temperature),
reagent
temperature can get close to
the set temperature more
quickly (see Fig. 2).
13
Sample Volume
Sample volume (ul)
10~200
14
Hotlid
Hot lid off
Off
Hot lid on (
℃
)
30~110