![Bioer BYQ6627E Скачать руководство пользователя страница 17](http://html1.mh-extra.com/html/bioer/byq6627e/byq6627e_operation-manual_2753692017.webp)
Gene Explorer Thermal Cyeler Operation Manual
Chapter Four Operation Instructions
-
11
-
2.2 Section temperature and constant temperature time
On the interface of PCR program setting, click the small icon on upper right of the screen,
and it will prompt the temperature and time setting box for setting temperature and constant
temperature time of the program section.
Run mode: Standard Mode and High-speed Mode.
Standard Mode running characteristics simulate BIOER TOUCH series.
2.3 Function of loop
On the interface of PCR program setting, click “
Loop
”, and it will prompt interface 1-1,
select the starting and ending sections of the loop, click “Determine” and it will prompt the
setting box 1-2 for loop times, input the times and click “OK”, the setting of loop section will
be finished.
Attention:
In one program, multiple loop sections can be set, but these sections cannot
be crossed, but only paralleled or included.