![Bioer BYQ6627E Скачать руководство пользователя страница 11](http://html1.mh-extra.com/html/bioer/byq6627e/byq6627e_operation-manual_2753692011.webp)
Gene Explorer Thermal Cyeler Operation Manual
Chapter Two Characteristics
―
5
-
Attention:
When the set value of hot lid temperature is lower than the set value of module
temperature, the hot lid temperature displayed on the screen will be higher than the
set value of hot lid temperature.
4 Software functions
(1) file edit (including file number, temperature and time of each section, setting of loop times, view,
revise and delete file, automatic file connection, gradient temperature setting; maximum number of
editable sections is 10, maximum number of steps is 100, maximum number of loop is 99, available
for secondary nesting, gradient temperature range is -6
℃
~6
℃
);
(2) file storage;
(3) file modification (including adjustment of ramp rate, automatic temperature & time
modification during cycling, time display and calibration);
(4) setting of temperature, time and loop times during file running, and display of instant values
(including voice prompt, estimation of overall and remaining time of file running, fault protection and
warning);
(5) stop file running (including pause file running, stop file running and automatic recovery after
outage).
Attention:
The above software functions are only for reference. Please understand that no
further notice will be given for any modification to the software functions.
5 Software version
Software version : V1.0