LIGHTWAVE 3 & 3+ SPECTROPHOTOMETERS USER’S MANUAL
22
USER INTERFACE
Step 1
Set the wavelength, 190 to 1100 nm, using the left
and right arrow or alphanumeric keys.
Step 2
Press the down arrow, set the number of standard
samples of known concentration, 1 to 9, using the
left and right arrow or alphanumeric keys.
Step 4
Press the down arrow, set the sample prompt,
“Yes” or “No”, to bring up the sample number option
before each measurement, using the left and right
arrow keys.
Step 3
Press the down arrow, enter up to 8-digits to define
the units that the primary result will be reported in.
OR
Open the units options using the options key:
Select one of the predefined units; “µg/ml”, “µg/µl”,
“pmol/µl”, “mg/dl”, “mmol/l”, “µmol/l”, “g/l”, “mg/l”,
“µg/l”, “U/l”, “%”, “ppm”, “ppb”, or “conc”, using the
left and right arrow keys.
Press the down arrow, select the number of decimal
places to display the concentration in, “Auto”, “0”,
“1”, or “2”, using the left and right arrow keys.
Implement the changes and return to the parameters
screen by selecting “OK” using the confirm key.
OR
Reject the changes and return to the parameters
screen by selecting “Cancel” using the escape key.