3. If the control solution testing result is out of range :
DETAILED INFORMATION
The meter provides you with plasma equivalent results.
Source: American Diabetes Association (2014). Clinical Practice
Recommendations. Diabetes Care, 37 (Supplement 1) : S16.
Please consult your doctor to determine a target range that works best for
you.
POSSIBLE CAUSE WHAT TO DO
Time of day Normal plasma glucose range for
people without diabetes (mg/dL)
Error in performing the test.
Control solution vial was
poorly shaken.
Expired or contaminated
control solution.
Control solution that is too
warm or too cold.
Defective test strip.
Meter malfunction.
Improper working of meter
and test strip.
Fasting and before meal
2 hours after meals
< 100 mg/dL (5.6 mmol/L)
< 140 mg/dL (7.8 mmol/L)
Read instructions thoroughly and
repeat the test again.
Shake the control solution
vigorously and repeat the test again.
Check the expiry date of the control
solution.
Control solution, meter, and test strips
should be at room temperature 20°C to
25°C (68°F to 77°F) before testing.
Repeat the test with a new test strip.
Please contact customer services.
Please contact customer services.