9 Product Maintenance
9.1 Cleaning Agent
Specifications of cleaning agent:
Cleaning Agent
Specification
Water
Distilled or deionized water
Alcohol
75% ethanol
Warning:
The cleaning agent Should be consistent with this instruction and the high
concentration may dissolve the carrier and the equipment surface coatings.
Do not use the hypochlorite-containing disinfectant (bleach liquid and chlorine)
or bleach to clean the instrument and its components.
Do not use open fire or do not clean the instrument near machines which prone to
sparks, and do not use heat fan to dry it, because it may cause misfire.
9.2 Daily Maintenance
Check both the waste liquid box and the waste pipette tip bag are empty before
the experiment
,
daily.If there is no emptying,please dispose the waste liquid and
the waste pipette tip to the designated place according to the laboratory
requirements,and conduct the necessary disinfection treatment.
Check the instrument countertop after the daily experiment, and keep it clean,
with no residual liquid and stains.If there is liquid residue, remove it with suction
paper and wipe the top with a clean soft cloth.
After the experiment, if the pipette tips are not used up, put them back to the
original box and seal them up
,
daily.
After the experiment,clean up the empty tip carrier shelf,reagent position,
samples,etc.,daily.
After the experiment,the operator needs to turn off the operation switch button
and the main power switch button of the instrument,daily.
When the instrument is not used for a long time, cover it with a dust cover and
cut off the power.
LCD screen can´t be wiped with water, just wipe it gently with clean soft cloth or
soft paper.
Do not move and collide the instrument after the installation and debugging.
Do not spill the solution into the instrument.
When the instrument is not used for a long time, turn off the power supply and