34
7 Software Update
7.1
Single Step Movement
Click [Instrument Management] and click "Sampling motor debugging" on the
instrument management interface. Under this interface, you can control the single step
movement of the adding sample arm/twisting cap arm.
[Sampling
arm]:After
selecting
the
target
board
position
at
"
", select the row and column, click" Move "to add sample arm to
move to the specified position, click" Take Tip"to take the tip, click"Drop Tip "to
remove the tip, and click" Init "to add sample arm to return to initialization and initial
zero position.
Figure 7-1
Screw-top arm movement: The sample carrier marked row and column number, find
the target location and input them, select the volume of the tes tube, click on the
[move] to the target location, click the [Open cover] to open, click [Close cover] the
tube cap will be buckled, click on the [Init] early screw-top arm will return to the
initial return to initialize the zero.
Figure 7-2
After selecting the specified position according to the above [Sampling arm/Cap
screwing arm], select the correct tip head and test tube, select the target position,
select the corresponding liquid parameters, click [Send aspiration parm], click [Probe
liquid] for liquid probe, click [Suction liquid ] for liquid suction, and the Z-axis
returns to zero after the liquid suction is completed. After selecting the target position
of liquid injection, click [Send injection Parameters] and click [Liquid suction] for
liquid injection.