5.2 Equipment alarm and handling measures
Alarm
prompt
Reaso
n
Actio
n
There is no tip in
the current hole
position, but there
is a tip in the rear
face position.
Click "Retry."
All tips are used up.
Please install a
new tip and click
"Retry".
The dosing arm is
moving out of step
in X or Y direction..
Click on "Init
Robot".
The dosing arm isis
moving out of step
in Z direction.
Click "Init Retry."
Insufficient
sample, reagent, or
control solution.
After adding an
appropriate
amount of
specimen,
reagent or
control solution,
click "Retry" or
"ReGetTips
Retry".
Aspiration does not
move to the lowest
position.
Click on "Asp
to lowest pos".
Dosing arm is
movingt out of step.
Click on "Init
Robot".
The gripper did not
grab the deep well
plate.
Check to see if
there is a deep
well plate at the
position of the
gripper. If there
is, click "Retry".
If not, add the
microplate
first
and then click
"Retry".
5.3 Handling when the instrument is deactivated
1.
Disconnect the power cable from the device.
2.
Close the instrument compartment door to prevent the equipment from contaminating the
dust.
Содержание BK-AutoHS96
Страница 1: ...Automatic Nucleic Acid Extraction System BK AutoHS96 User Manual BIOBASE GROUP Version 2020 07...
Страница 18: ...17 10 Rear service panel For mechanical arm maintenance...
Страница 20: ...19 Figure 2 3 Internal Structure of Automatic Nucleic acid extraction system BK AutoHS96...
Страница 23: ...22 7 Waste tip bag Collect abandoned tips...
Страница 27: ...26 numbering function it is used to define the number of the first sample...
Страница 52: ...51 Acid Extraction System or individual parts are...
Страница 54: ...53 Appendix Parts List No Component name Specifications Quantity 1 Fuse 6 3A 2...