iv
Safe Use Specifications and Compliance
Table 5 lists the safe use specifications for the CFX96 system or the CFX384 System. Shielded cables
(supplied) must be used with this unit to ensure compliance with the Class A FCC limits.
R
EGULATORY
C
OMPLIANCE
This instrument has been tested, and found to be in compliance with all applicable requirements of the
following safety and electromagnetic standards:
• IEC 61010-1:2001 (2nd Ed.), EN61010-1:2001 (2nd Ed). Electrical Equipment For
Measurement, Control, and Laboratory Use - Part 1: General Requirements
• IEC 61010-2-010:2005, EN61010-2-010:2003. Safety requirements for electrical equipment
for measurement, control and laboratory use. Part 2-010: Particular requirements for
laboratory equipment for the heating of materials
• IEC 61010-2-081:2001+A1, EN61010-2-081:2002+A1. Safety requirements for electrical
equipment for measurement, control and laboratory use. Part 2-081: Particular requirements
for automatic and semi-automatic laboratory equipment for analysis and other purposes
(includes Amendment 1)
• EN 61326-1:2006 (Class A) Electrical equipment for measurement, control and laboratory
use. EMC requirements, Part 1: General requirements
This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in
accordance with the instruction manual, may cause harmful interference to radio communications.
Operation of this equipment in a residential area is likely to cause harmful interference in which case the
user will be required to correct the interference at his own expense.
Table 5. Safe Use Specifications
Safe Use Requirements
Specifications
Temperature
Indoor use
Ambient temperature of 15—31
o
C. Relative humidity
maximum of 80% noncondensing
Altitude
Up to 2,000 meters above sea level