CFX96 and CFX384 Systems Manual
87
RFU Spreadsheet
Select the
RFU
spreadsheet to see the RFU readings for each well acquired at each cycle of
the experiment. Select individual fluorophores by clicking a tab at the bottom of the
spreadsheet. The well number appears at the top of each column, and the cycle number
appears to the left of each row (Figure 87).
Figure 87. RFU spreadsheet in the Quantitation Data tab.
The RFU spreadsheet includes the type of information shown in Table 28.
Melt Curve Tab
For DNA-binding dyes and noncleavable hybridization probes, the fluorescence is brightest
when the two strands of DNA anneal. Therefore, as the temperature rises towards the melting
temperature (Tm), fluorescence decreases at a constant rate (constant slope). At the Tm, there
is a dramatic reduction in the fluorescence with a noticeable change in slope. The rate of this
change is determined by plotting the negative first Regression of fluorescence versus
temperature (-d(RFU)/dT). The greatest rate of change in fluorescence results in visible peaks
and represents the Tm of the double-stranded DNA complexes.
The software plots the RFU data collected during a melt curve as a function of temperature. To
analyze melt peak data, the software assigns a beginning and ending temperature to each
peak by moving the threshold bar. The floor of the peak area is specified by the position of the
melt threshold bar. A valid peak must have a minimum height relative to the distance between
the threshold bar and the height of the highest peak.
Open the Melt Curve tab (Figure 88) to determine the melting temperature (Tm) of amplified
PCR products. This tab shows the melt curve data in these four views:
•
Melt Curve.
View the real-time data for each fluorophore as RFUs per temperature for
each well
•
Melt Peak.
View the negative regression of the RFU data per temperature for each well
•
Well Selector.
Select wells to show or hide the data
•
Peak spreadsheet.
View a spreadsheet of the data collected in the selected well
NOTE: This spreadsheet only shows as many as two peaks for each trace. To see
more peaks, click the
Melt Curve Data
Table 28. RFU spreadsheet contents
Information
Description
Well number
(A2, A3, A4, A5, A6...)
Well data, listed by position in the plate for all the loaded wells
Cycle
One round of denaturation, annealing, and extension, or one
round of annealing and extension steps in a protocol