11.2 Resetting the notification or alarm messages
The “RESET” button, which serves to acknowledge and reset the indication, will become visible automat-
ically whenever a notification or an alarm message appears.
1.
Remove the cause of the disturbance or wait until the unit compensates for the reason of the error.
2.
Press the “RESET” button to reset the notification or alarm message.
CAUTION
In case the “RESET” button does not cancel the notification or alarm indication, the reason
for the disturbance was not removed correctly.
Contact BINDER service.
The “RESET” button allows resetting the notification or alarm messages regarding tempera-
ture and humidity only within a tolerance sector of +/- 2 °C resp. +/- 5 % r.H.
With values outside this range, contact BINDER service.
If the humidity deviates by more than +/- 5 % r.H. from the set-point, this triggers an alarm
message.
In order to avoid such limit alarms when operating without humidity (humidity switch (3) OFF):
•
in Manual Mode set the humidity set-point to 0 % r.H.
•
in Program Mode enter a humidity sub-program with the humidity set-points set to 0 % r.H.