KBF-S (E6) 08/2018
page 30/88
5.2
Performance during and after power failures
During a power failure, all controller functions are shut down.
After the power returns, all functions return to the same status the chamber had before power failure. The
set-points are immediately resumed.
If during power failure an alarm has occurred (tolerance range, safety controller etc.), confirm the alarm.
See chap. 13.
6. Start up
After connecting the power supply and filling the freshwater can, turn on the chamber by its main power
switch (1). The lit pilot lamp shows the chamber is ready for operation.
The controller shows normal display and controls temperature and humidity to the last entered values.
For operation with humidity, the humidifying and dehumidifying system must be activated (chap. 8).
For operation with humidity:
•
Activate the humidifying and dehumidifying system (chap. 8).
•
Open the freshwater can tap (chap. 4.1.3)
•
Do not firmly close the lid of the freshwater can to allow air to enter the freshwater can. As a result,
obstruction of the water supply due to negative pressure inside the can is avoided.
After the first turning on of the chamber or after an interruption of the power supply the relative humidity
will increase after a delay of about 20 minutes. During this period, the relative humidity can drop consid-
erably.
Warming chambers may release odors in the first few days after commissioning. This is not a quality
defect. To reduce odors quickly we recommend heating up the chamber to its nominal temperature for
one day and in a well-ventilated location.
WARNING: If customer should use a BINDER chamber running in non-supervised continu-
ous operation, we strongly recommend in case of inclusion of irrecoverable specimen or
samples to split such specimen or samples and store them in at least two chambers, if this is
feasible.
If the function “Language selection at restart” has been activated (chap. 11.5, factory setting ON), the
following settings are checked upon start up:
•
Menu language
(chap. 11.1):
Use the
arrow buttons
to select the desired language, confirm with the
OK button
•
Temperature unit
(chap. 11.2):
Use the
arrow buttons
to select the desired temperature unit, confirm with the
OK button.
•
Current date
(chap. 11.3), format DD MM YYYY:
Use the
arrow buttons
to set the day, continue with the
OK button
.
Use the
arrow buttons
to set the month, continue with the
OK button
.
Use the
arrow buttons
to set the year, confirm with the
OK button
•
Current time
(chap. 11.4), format HH:MM:
Use the
arrow buttons
to set the hours, continue with the
OK button
.
Use the
arrow buttons
to set the minutes, confirm with the
OK button
Содержание KBF-S 240
Страница 75: ...KBF S E6 08 2018 page 75 88 24 7 Dimensions Dimensions size 240...
Страница 76: ...KBF S E6 08 2018 page 76 88 Dimensions size 720...
Страница 77: ...KBF S E6 08 2018 page 77 88 25 Certificates and declarations of conformity 25 1 EU Declaration of Conformity...
Страница 78: ...KBF S E6 08 2018 page 78 88...
Страница 79: ...KBF S E6 08 2018 page 79 88...
Страница 81: ...KBF S E6 08 2018 page 81 88...
Страница 82: ...KBF S E6 08 2018 page 82 88 26 Product registration...
Страница 89: ......