KBF-S (E6) 08/2018
page 41/88
11.6 Setting the chamber address
This setting is required for the communication with the BINDER communication software APT-COM 3
DataControlSystem. The chamber address settings in the chamber controller and in the software must be
identical.
Required access level: “Admin”.
Path:
Normal display
Settings
Chamber
Chamber
address
Press the
OK button
to enable the setting.
Setting the chamber address
The current setting flashes. Enter the desired address with the
arrow
buttons
. Setting range: 1 up to 254
Confirm the entry with the
OK button
.
Chamber address
With the
arrow-down button
you can now change to the next parameter (display brightness).
With the
Back button
you can go back to the “
Chamber
” submenu and, repeatedly pressing it, to
Nor-
mal display
.
11.7 Display brightness
Required access level: “Admin”.
Path:
Normal display
Settings
Chamber
Brightness
Press the
OK button
to enable the setting.
Setting the display brightness
The current setting flashes. Enter the desired value with the
arrow
buttons
. Setting range: 10% up to 100%
Confirm the entry with the
OK button
.
Brightness
With the
arrow-down button
you can now change to the next parameter (audible alarm, chap. 13.2).
With the
Back button
you can go back to the “
Chamber
” submenu and, repeatedly pressing it, to
Nor-
mal display
.
Содержание KBF-S 240
Страница 75: ...KBF S E6 08 2018 page 75 88 24 7 Dimensions Dimensions size 240...
Страница 76: ...KBF S E6 08 2018 page 76 88 Dimensions size 720...
Страница 77: ...KBF S E6 08 2018 page 77 88 25 Certificates and declarations of conformity 25 1 EU Declaration of Conformity...
Страница 78: ...KBF S E6 08 2018 page 78 88...
Страница 79: ...KBF S E6 08 2018 page 79 88...
Страница 81: ...KBF S E6 08 2018 page 81 88...
Страница 82: ...KBF S E6 08 2018 page 82 88 26 Product registration...
Страница 89: ......