3. Set the number of cycles to infinite in the program start level and start the program
The described example program must be started once at the precise moment of temperature
change (on Monday e.g., at 0.01 or at 7.00). If the program cannot be manually started at the
desired moment, you can program a suitable program delay-time of 99 hs. 59 min. max. After
rundown of this delay time, the program will start automatically (chap. 9).
12.3 Proceeding in detail
1. Settings in the user level:
•
Setting the safety controller to “Limit” and 3 °C above the maximum temperature value of the
program
The temperature set-point of the safety controller (over temperature protection class 3.1) is shown and
can be changed. An absolute value in case of setting “limit”, or a relative value in case of setting
“Offset” can be entered Select setting “Limit” and enter a value, which is by 3 °C above the maximum
value (i.e., 23 °C /
73.4°F
).
•
Setting the maximum section duration (Prg. Time) to 999 hs. 59 min.
The maximum length of a program section can be set – in common for all program sections – to 99 hs
59 min or to 999 hs 59 min. Select setting 999:59.
When changing the maximum duration setting, pre-existing programs will be deleted in
the program editor.
•
Switching off the tolerance limits function
Select setting “0” meaning tolerance limits off. Thus an interruption of the program course during the
heating-up or cooling-down phases during the rapid “set-point step” phase is avoided.
•
Deactivating the week program timer
Deactivate the week program timer before entering a program (factory setting). Otherwise, any setting
of the operation lines in the program editor is ineffective.
Normal Display
Display 1 shows
e.g. 19.8
(actual temperature value)
Display 2 shows
e.g. 15.05.06 13:52 - -
(actual date and time, actual switching state of week
program timer channel 1: Off, channel 2: Off)
Press down key
for 5 sec
Display 1 shows
e.g. 0000
Display 2 shows
PROGRAM EDITOR
(you are in the program editor)
Press down key
for 5 sec
Display 1 shows
0000
Menu visible only if week program timer is activated
Display 2 shows WEEK PROG. EDITOR
(you are in the week program editor)
Press down key
for 5 sec
Display 1 shows
0000
Display 2 shows
USER-LEVEL
(you are in the user level)
Press program key
Содержание KB 23
Страница 75: ...KB 23 KB 23 UL E3 1 02 2019 page 75 87 20 7 KB KB UL 23 dimensions...
Страница 77: ...KB 23 KB 23 UL E3 1 02 2019 page 77 87...
Страница 78: ...KB 23 KB 23 UL E3 1 02 2019 page 78 87...
Страница 80: ...KB 23 KB 23 UL E3 1 02 2019 page 80 87...
Страница 81: ...KB 23 KB 23 UL E3 1 02 2019 page 81 87 22 Product registration...