KB 23 / KB 23-UL (E3.1) 02/2019
page 17/87
Do not install or operate the chamber in potentially explosive areas.
DANGER
Explosion hazard.
Danger of death.
∅
Do NOT operate the chamber in potentially explosive areas.
KEEP explosive dust or air-solvent mixtures AWAY from the vicinity of the chamber.
4.
Electrical connection
The chambers are supplied ready for connection. They come with a fixed power connection cable of at
least 1800 mm /
70.87 in
in length. They are equipped with a semi time-lag fine-wire fuse.
Model
Power plug
Nominal voltage
±
10% at
the indicated power
frequency
Current
type
Chamber
fuse
KB 23 (E3.1)
Shock-proof plug
230 V at 50 Hz
1N~
10 A
KB 23-UL (E3.1)
NEMA 5-15
115 V at 60 Hz
1N~
12.5 A
•
The domestic socket must also provide a protective conductor. Make sure that the connection of the
protective conductor of the domestic installations to the chamber’s protective conductor meets the
latest technology. The protective conductors of the socket and plug must be compatible!
•
Prior to connection and start-up, check the power supply voltage. Compare the values to the specified
data located on the chamber’s type plate (left chamber side, bottom right-hand, see chap. 1.4).
•
When connecting, please observe the regulations specified by the local electricity supply company as
well as the VDE directives (for Germany). We recommend the use of a residual current circuit breaker.
•
Pollution degree (acc. to IEC 61010-1): 2
•
Over-voltage category (acc. to IEC 61010-1): II
CAUTION
Danger of incorrect power supply voltage.
Damage to the equipment.
Check the power supply voltage before connection and start-up.
Compare the power supply voltage with the data indicated on the type plate.
See also electrical data (chap. 20.4).
To completely separate the chamber from the power supply, you must disconnect the power
plug. Install the chamber in a way that the power plug is easily accessible and can be easily
pulled in case of danger.
Содержание KB 23
Страница 75: ...KB 23 KB 23 UL E3 1 02 2019 page 75 87 20 7 KB KB UL 23 dimensions...
Страница 77: ...KB 23 KB 23 UL E3 1 02 2019 page 77 87...
Страница 78: ...KB 23 KB 23 UL E3 1 02 2019 page 78 87...
Страница 80: ...KB 23 KB 23 UL E3 1 02 2019 page 80 87...
Страница 81: ...KB 23 KB 23 UL E3 1 02 2019 page 81 87 22 Product registration...