KB 23 / KB 23-UL (E3.1) 02/2019
page 26/87
Program entry as set-point ramp (example):
W/°C
t/min.
S01
S02
S03
S04
0
20
40
60
80
100
30
120
180
380
S05
Program table corresponding to the diagram (with default setting “Ramp”):
Section
SEC
Temperature
set-point
[ °C]
TEMP
Section length
[hh.mm]
TIME
Fan speed
[%]
FAN
Operation lines
*
O.LINE
S01
40
00:30
50
000
S02
60
01:30
100
000
S03
90
01:00
100
000
S04
90
03:20
100
000
S05
20
00:01
100
000
*
Only with option zero-voltage relay outputs via operation lines, see chap. 16.4.
The values of such a program table can now be entered to the RD3 program controller (chap. 8.2).
Program entry as set-point step (example):
W/°C
t/min.
S01 S02 S03 S04 S05 S06 S07 S08
0
20
40
60
80
100
30
120
180
380