KB (E3.1) 08/2012
page 36/84
Deactivate the week program timer (factory setting, setting in the user level, chap. 10) before
starting a program.
Step 1 – Program selection (only with program type “2 programs” set):
Normal Display
Hit program key
Display 1 shows
e.g. 1
(actually selected program)
Display 2 shows
SEL.PRG.
(select program 1 or 2)
Enter program number 1 or 2 with arrow
keys
Value is shown in display 1.
Next step
–
entry of program course settings
Hit program key
Display 1 shows
e.g. 00.00
(entered delay time hh.mm)
Display 2 shows
RUN TIME
(enter delay time of program start)
Set delay time in hh.mm with arrow keys
Value is shown in display 1.
Hit program key
Display 1 shows
e.g. -1
(actually selected number of program cycles)
Display 2 shows
REPEAT
(enter number of program cycles)
Select number of cycles -1, 0, 1 etc. with
arrow keys
Value is shown in display 1.
Last step – program start:
Hit program key
Display 1 shows
e.g. 1
(selected program)
Display 2 shows
RUN PRG.
(Question: start selected program?)
Hit program key
Display 1 shows
e.g. 25.5
C
actual temperature value
Display 2 shows
P01:S01 01:49:39
(time running backwards)
(actual program P01, actual section S01, and remaining
time of program section S01)
Program is running. The green LED (3d) lights up.
In addition to the green LED (3d) indicating a running program, the LED (3a) is lit if the heating is active,
or LED (3b) if the refrigeration is active, or none of both if the actual temperature equals the set-point.
During program course the arrow keys and the EXIT button are not functional.
By pressing the program key
for 3 seconds, you can terminate the program course.
Содержание 9020-0112
Страница 5: ...KB E3 1 08 2012 page 5 84 Product registration...
Страница 85: ......