BD / ED / FD (E2) 01/2022
page 13/66
•
Continued operation of the chamber during an obvious malfunction
•
Insertion of objects, particularly metallic objects, in louvers or other openings or slots on the chamber
•
Human error (e.g. insufficient experience, qualification, stress, exhaustion, laziness)
To prevent these and other risks from incorrect operation, the operator shall issue operating instructions.
Standard operating procedures (SOPs) are recommended.
1.10 Residual Risks
The unavoidable design features of a chamber, as well as its proper field of application, can also pose risks,
even during correct operation. These residual risks include hazards which, despite the inherently safe
design, existing technical protective equipment, safety precautions and supplementary protective
measures, cannot be ruled out.
Messages on the chamber and in the Operating Manual warn of residual risks. The consequences of these
residual risks and the measures required to prevent them are listed in the Operating Manual. Moreover, the
operator must take measures to minimize hazards from unavoidable residual risks. This includes, in
particular, issuing operating instructions.
The following list summarizes the hazards against which this Operating Manual and the Service Manual
warn, and specifies protective measures at the appropriate spots:
Unpacking, Transport, Installation
•
Sliding or tilting the chamber
•
Setup of the chamber in unauthorized areas
•
Installation of a damaged chamber
•
Installation of a chamber with damaged power cord
•
Inappropriate site of installation
•
Missing protective conductor connection
Normal operation
•
Assembly errors
•
Contact with hot surfaces on the housing
•
Contact with hot surfaces in the interior and inside of doors
•
Emission of non-ionizing radiation from electrical operating resources
•
Contact with live parts in normal state
Cleaning and Decontamination
•
Penetration of water into the chamber
•
Inappropriate cleaning and decontamination agents
•
Enclosure of persons in the interior
Malfunction and Damage
•
Continued operation of the chamber during an obvious malfunction or outage of the heating system
•
Contact with live parts during error status
•
Operation of a unit with damaged power cord
Maintenance
•
Maintenance work on live parts.
•
Execution of maintenance work by untrained/insufficiently qualified personnel
•
Electrical safety analysis during annual maintenance not performed
Содержание 9010-0073
Страница 54: ...BD ED FD E2 01 2022 page 54 66...
Страница 55: ...BD ED FD E2 01 2022 page 55 66 13 2 EU Declaration of Conformity for ED...
Страница 56: ...BD ED FD E2 01 2022 page 56 66...
Страница 57: ...BD ED FD E2 01 2022 page 57 66 13 3 EU Declaration of Conformity for FD...
Страница 58: ...BD ED FD E2 01 2022 page 58 66...
Страница 60: ...BD ED FD E2 01 2022 page 60 66...
Страница 67: ......