86
ADVANCED USE MANUAL DD700
3.8 Weight repeater
The paragraphs below illustrate the parameters that need to be set for the repeater scale. For the
Customisation menu and for Test procedures refer to the dedicated paragraphs.
3.8.1
Setup Menu/Repeater scale/Repeater scale parameters/Serial port/xy com
In order to establish communication between main terminal and repeater, the settings of the
parameters listed below must be the same as those on the main terminal.
Baud rate
600, 1200, 2400, 4800, 9600, 19200, 38400,57600,115200
Number of bits
7.8
Number of stop
bits
1.2
Parity
Even, Odd, None
Communication
interface
RS232, RS422, RS485
Serial com
control
Hardware, Software, None
Hardware: CTS and RTS signals are used. The input CTS signal must be placed
on +12V to enable transmission on the TX feet. The terminal places the output
RTS signal on +12V to report its availability to receive (only available on COM1).
Software: sending the character
<XON>(11H)
enable transmission;
<XOFF>(13H)
suspend transmission;