D440IS
2-105
11.7.4 Terminal Tests/Keypad
Follow the instructions given on the display.
11.7.5 Terminal Tests/Terminal Configuration Report
If a printer port has been enabled through the relative menu, the terminal
configuration report will be printed via this port. If no port has been
enabled, connect a printer and a PC to the COM1 serial port with fixed
9600,8,N,1 configuration.
The overall setup of the terminal will be transmitted to it.
11.7.6 Terminal Tests/Analogue output
To calibrate the powered output, connect an mA capacity tester to the
terminals. Regulate the output with the keys indicated on the display
until there is a 4.00 mA current reading for the zero mark and 20.00 mA
for the full scale.
Check the setting using a load cell simulator.
Содержание D440IS
Страница 1: ...Terminal D440IS Advanced user manual Code 81320155 EDITION 08 06 2015 ...
Страница 2: ......
Страница 114: ......