background image

5. How to use

5-1 Leveling

(1) Clean the bottom surface of Level Master and Machine table. (Scratch and dent mark, oil, dust etc)

(2) Place the Level master on the base level. X & Y axis need to be carefully adjusted.

(3) Turn on the power of Level Master. 

(4) When the power is turned on, High mode is the initial setting. (Inclination: 0.01mm-0.10mm)

If LED (

RED

) is blinking at 0.08mm, the inclination is over 0.1mm.

And you need to change the Mode from High to Low. (0.1mm-1mm)

If LED (

RED

) is on or other LED is on, the inclination is within 0.08mm,  follow the procedure (7) below in High mode.

Change Mode from High to Low

When Level Master is turned on,

LED (

RED

) blinks

Содержание LVM-01

Страница 1: ...LED RED blinking over 0 8mm Inclination 1m LED RED blinking over 0 08mm Inclination 1m LED RED on within 0 8mm Inclination 1m LED RED on within 0 08mm Inclination 1m LED ORANGE blinking within 0 6mm...

Страница 2: ...e 0 08mm 0 01mm Inclination 1m 10 Buzzer When center position LED is on buzzer makes sound 5 Level Indication Y axis In Low mode 0 8mm 0 1mm Inclination 1m In High mode 0 08mm 0 01mm Inclination 1m 6...

Страница 3: ...cuit Cover Cover convex Set screw Battery box Battery LR03 AAA x4pcs Loosen the screw on the battery box cover Set 4 batteries LR03 AAA battery as described on the bottom of the battery box After sett...

Страница 4: ...always stay in the center 0 position Make sure to calibrate the unit depending on the environment to be used Place the unit in the center of base level Before placing the unit on the base level make s...

Страница 5: ...tion 1 of X Y levels is out of 0 08 micron Red LED is blinking See Page 6 How to check levels in Low Mode Case 1 Case 2 How to check levels in High Mode Rotate 180 degrees Rotate 180 degrees Red LED i...

Страница 6: ...Adjust the levels so that both of X Y levels stay within 0 8 micron See Page 7 How to calibrate in Low Mode How to calibrate in Low Mode Change the Mode to Low Mode After the mode is changed 1 of X Y...

Страница 7: ...s start to blink Rotate the Level Master for 180 degrees referring to the markings on base level Misalignment and insufficient rotation less than 180 degrees causes the error in calibration process No...

Страница 8: ...gain Buzzer sounds Pi and calibration is completed Adjust levels of X Y axis of base level within 0 1 micron If one of 2 levels is adjusted within 0 1 micron Blue LED is on and buzzer sounds Pi Pi Whe...

Страница 9: ...start to blink Rotate the Level Master for 180 degrees referring to the markings on base level Misalignment and insufficient rotation less than 180 degrees causes the error in calibration process Not...

Страница 10: ...in High Mode When the actual level is over 0 08 micron Even if you push the Set switch after rotating 180 degrees to complete the calibration procedure 4 orange LEDs on 0 6 micron position in Low Mode...

Страница 11: ...power of Level Master 4 When the power is turned on High mode is the initial setting Inclination 0 01mm 0 10mm If LED RED is blinking at 0 08mm the inclination is over 0 1mm And you need to change the...

Страница 12: ...1mm When only 1 of 2 axis X Y is adjusted until inclination within 0 1 LED BLUE blinks and buzzer makes sound like Pi Pi When both of 2 axis X Y is adjusted until 0 1 inclination LED BLUE lights on a...

Страница 13: ...of 2 axis X Y is adjusted until inclination within 0 01 LED BLUE blinks and buzzer makes sound like Pi Pi When both of 2 axis X Y is adjusted until 0 01 inclination LED BLUE lights on and the buzzer m...

Страница 14: ...g You can turn on off the buzzer Keep pushing the Set switch 5 3 Default setting If you want to return to the default setting keep pushing the Set switch and turn on the Power Data inside Level Master...

Страница 15: ...liquid Do not leave Level master on the machine table surface flat and jig Do not make the disassembling and modification at customer It inferiors the performance and life of Level Master Remove the...