③
Push the [Set switch] once again. Buzzer sounds
[Pi: ] and calibration is completed.
↓
Now please start leveling.
In case of [Calibration] in Low Mode:
*When the actual level is over 0.8 micron,
In case of [Calibration] in High Mode:
*When the actual level is over 0.08 micron,
Even if you push the [Set switch] after rotating 180 degrees to complete the
calibration procedure, 4 orange LEDs (on 0.6 micron position in Low Mode/ on
0.06 micron position in High Mode) start to blink and buzzer sounds [Pi, Pi, Pi, Pi]
to inform you that you can not complete the procedure.
4 orange LEDs start to blink
Push the [Set switch]
Buzzer sounds
[Pi: ]