6
Conducted
disturbances
induced by RF
fields
IEC 61000-4-6
3 V
RMS
0,15 MHz – 80 MHz
6 V
RMS
in ISM bands
0,15 MHz – 80 MHz
80% AM at 1 kHz
3 V
RMS
0,15 MHz – 80 MHz
6 V
RMS
in ISM and
amateur bands
0,15 MHz – 80 MHz
80% AM at 1 kHz
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site
survey
1
should be less than the
compliance level in each
frequency range.
Interference may occur in the
vicinity of equipment marked
with the following symbol:
Radiated RF
EM fields
IEC 61000-4-3
?
3 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
3 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
Proximity
fields from RF
wireless
communicatio
ns equipment
IEC 61000-4-3
Test freq.
[MHz]
Max. power
[W]
Immunity
test level [V/
m]
Distance: 0.3 m
385
1.8
27
450
2
28
710, 745,
780
0.2
9
810, 870,
930
2
28
1720, 1845,
1970
2
28
2450
2
28
5240, 5500,
5785
0
9
NOTE: U
T
is the AC mains voltage prior to application of the test level.
Essential performance per IEC 60601-1: The essential performance is to maintain the visual
luminous intensity of the LED and the motor speed. The maximum speed deviation is ±5%.
1. Field strengths from fixed transmitters, such as base stations for radio (cellular/
cordless) telephones and mobile field radios, amateur radios, AM and FM radio
broadcasts and TV broadcasts cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fixed RF transmitters, an electro-
magnetic site survey should be considered. If the measured field strength in the lo-
cation in which the iOptima INT is used exceeds the RF compliance level mentioned
above, the iOptima INT should be observed to verify that it is operating normally. If
abnormal operation is observed, additional measures may be necessary, such as
reorienting or relocating the iOptima INT.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment -