Additional notes.
The Tecal 700 is a precision calibration instrument designed for optimum durability and trouble free
operation but it must be handled with care.
The unit should always be carried in the upright position to prevent the insert blocks from being dropped
out and damaged.
Test probes and the Tecal 700 block may have differing thermal expansion rates. The precision bores are
manufactured to a very close tolerance to allow for probe expansion inside the well as the block heats
up. Otherwise, the probe may become stuck in the well.
Some test probes have temperature limits so it is advisable not to exceed this temperature. Ensure that
the air temperature above the Tecal 700 does not exceed the probe handle’s temperature limit.
The air above the block may be over 200°C and if the probe handle limits are exceeded, the probe may
be permanently damaged.
unpacking
When unpacking please ensure that the following have been removed from the packaging:
• Tecal 700 Calibrator
• Calibration certificate
• Mains cable
• TechneWorks software on CD
• Guarantee card
The user is advised to keep the original packaging in case the instrument ever needs to be returned for
service or repair. Techne accepts no responsibility for damage incurred unless the unit is correctly packed
and transported in its original packaging.
3
Содержание Techne TECAL 700
Страница 1: ...TECAL 700 operator s manual Issue 2 10 15 ...