usEr opErATIon sET up
Place the unit on an even horizontal surface with clearance around the unit of 20cm on all sides and 1m
above. The area around the calibrator should be free of draught, dirt, flammable substances, etc.
Plug the mains cable with the correct supply into the unit.
Carefully lower the
slim metric inserts or slim Imperial inserts into the relevant unit.
Note that Tecal 700 has Slim Insert blocks to suit metric or imperial models.
Generally if the Tecal 700 has metric bores a metric insert should be used.
Similarly if the Tecal 700 has imperial bores then imperial inserts should be used.
The well should be clear of any dirt, foreign objects, grit or grease before the insert is fitted.
The Slim insert block sits proud of the block and has 2 removal holes for use with the insert extraction
tool.
Turn on the power to the unit with the switch above the mains cable inlet.
The unit will display the current temperature and heat or cool to the current set point. When the unit
has reached the set point temperature the display will stabilise and will be stable after 10 minutes at
temperature.
Detailed instruction on programming the Tecal 700 is given below.
21
switching on
When you first switch on, the display will
show the edition of the software which
your unit uses. For example software issue
“F 1.00” would be shown.
This will display for 1 second then the
actual block temperature will be indicated.
set point
When the
sET
button is pressed the set point
LED illuminates.
The front panel controls consist of four push
buttons and a five digit LED display.
set
Test
usEr opErATIon – dIspLAy InTErfACE funCTIons
70000
set
Test
Set Point
F 100