restore Factory settings
All changes to the instrument defaults and settings can be
removed and the unit returned to its factory settings.
●
From the Administrator menu touch
restore Factory
settings
.
●
Touch the
none
button and enter the administrator
password followed by
Back
.
●
Touch
restore settings
and the unit will be reset to
factory defaults.
●
Touch
cancel
to return to the Administrator menu
without making any changes.
●
Touch
Back
to return to the Settings menu.
Heated lid
Use this to set the heated lid temperature to approximately
10ºC higher than the hottest block temperature in the
program.
●
Touch
on/off
to toggle between settings.
●
Touch the temperature button to set the required lid
temperature (35 to 115ºC).
Preheat lid
Set to On to pre-heat the lid before the thermal cycling
program begins. The lid will pre-heat for 2 minutes before
the program starts.
●
Touch
on/off
to toggle between settings.
sample cooling
Set to On to cool the block to 4ºC while the lid preheats.
●
Touch
on/off
to toggle between settings.
●
Touch
more settings
for further program defaults or
Back
to return to the Settings menu.
32
ProGram DeFaults
The program default settings will apply to every new program created on the
3
Prime. However individual
programs can be edited as required.