Bibby Sterilin Stuart SI500 Скачать руководство пользователя страница 3

1

SHAKING INCUBATOR
SI500
Instructions for use

1. Introduction

Thank you for purchasing this piece of Stuart
equipment. To get the best performance from
the SI500 please read these instructions carefully
before use. Before discarding the packaging
check that all parts are present and correct.

For your own safety and that of others
please read and understand the safety
advice given below before using the
equipment.

2. Safety advice before use

If the equipment is not used in the manner
described in this manual and with

accessories other than those recommended by
Stuart, the protection provided might be
impaired.

This equipment is designed to operate under the
following conditions:

For indoor use only

Use in a well ventilated area

Ambient temperature range +5ºC to +40ºC

Altitude to 2000m

Relative humidity not exceeding 80%

Mains supply fluctuation not exceeding 10%

Over-voltage category II IEC60364-4-443

Pollution degree 2

Use with a minimum distance all around of
200mm from walls or other items

Not to be used to heat or mix materials which
could lead to liberation of gases, explosions,
implosions

Not to be (used or) placed in areas known to
be hazardous

Do not attempt to lift or carry the unit alone.

Never move or carry the unit when in use or
connected to the mains electricity supply.

Do not use in a hazardous atmosphere or with
hazardous materials.

Ensure that the load is balanced.

Do not attempt to stop movement by hand.

In the case of mains interruption, the unit will
not restart on restoration of the electricity supply.

In the case of mechanical interruption, (e.g.
motor stall), the unit will continue operating on
removal of interruption.

Mechanical energy can lead to breakage of glass
vessels. Use with care.

3. General description

The Stuart SI500 orbital incubator is ideal for
mixing and aerating various biological samples
under accurate temperature controlled
conditions. The rotating platform has a digital
speed setting providing a smooth, uniform
circular motion with an orbit of 16mm. The
speed range is 30 to 300rpm. There is also an
adjustable digital countdown timer that
automatically stops after the set time (1 second
steps to 1h, 1 minute steps to 24h or 1h steps to
9 days). The shaking platform has a drawer
mechanism for easy access to the incubation
vessels. The maximum load is 10kg. The
temperature is controlled by a simple “press to
set” system in 0.1ºC intervals from ambient
+5ºC to 60ºC.

The SI500 comes fitted with a fully adjustable
cradle system to accommodate a wide range of
flasks and bottles (maximum size 1l flask).
Optional accessories include racks to
accommodate 1.5ml, 15ml and 50ml tubes.

4. Preparation for use

4.1 Electrical installation

THIS INSTRUMENT MUST BE EARTHED

Before connecting the instrument please read
and understand these instructions and ensure
that the line supply corresponds to that shown
on the rating plate. The instrument is designed
for use on 230V 50Hz.The power consumption
of the unit is 300W. The instrument is fitted with
an IEC socket at the rear of the instrument for
connection of the mains lead.

Содержание Stuart SI500

Страница 1: ...t stop mode rpm time on off press to start temperature C on off orbital incubator SI500 Shaking Incubator SI500 Instructions for use Mode d emploi Istruzioni per uso Instrucciones de funcionamiento Ge...

Страница 2: ...rbital incubator SI500 start stop mode rpm time on off press to start temperature C on off orbital incubator SI500 English Temperature control panel Speed control panel USB connector Locating holes fo...

Страница 3: ...In the case of mains interruption the unit will not restart on restoration of the electricity supply In the case of mechanical interruption e g motor stall the unit will continue operating on removal...

Страница 4: ...ectricity supply DO NOT SWITCH ON 4 3 Place the unit on a firm level non slip surface ensuring that there is sufficient free space on all sides without coming into contact with anything during use 2 4...

Страница 5: ...The minimum set temperature is ambient 5 C and the maximum 60 C 3 5 4 USB Connectivity Your SI500 has capability to connect to a PC via USB for long term tracking of the incubator temperature whilst...

Страница 6: ...installation adaptation modification fitting of non approved parts or repair by unauthorised personnel Bibby Scientific Ltd Stone Staffordshire ST15 0SA United Kingdom Tel 44 0 8449 360234 Fax 44 0 8...

Страница 7: ...stop mode rpm time on off press to start temperature C on off orbital incubator SI500 Panneau de commande de la temp rature Panneau de commande de la vitesse Connecteur USB Orifice de positionnement...

Страница 8: ...s utiliser dans une atmosph re dangereuse ou en pr sence de substances dangereuses S assurer de l quilibre de la charge Ne pas tenter de stopper le mouvement de l appareil manuellement En cas de coupu...

Страница 9: ...a plaque signal tique de l appareil afin de vous assurer que la fiche et les fusibles sont adapt s la tension et la puissance mentionn es Les couleurs des fils du cordon d alimentation sont les suivan...

Страница 10: ...nutes et enfin en jours et en heures jusqu un maximum de 9 jours Lorsque l cran affiche la dur e souhait e appuyez sur le bouton start stop marche arr t afin de mettre l appareil en marche Il est poss...

Страница 11: ...les des pi ces d tach es fournies ou sp cifi es par Stuart ou l un de ses repr sentants doivent tre utilis es L installation de pi ces non approuv es peut affecter l efficacit des dispositifs de s cur...

Страница 12: ...deur 408 mm Largeur 422 mm Hauteur maximale des r cipients 250 mm Dimensions de la plate forme 335 x 335 mm Caract ristiques techniques Temp rature De la temp rature ambiante 5 C 60 C R solution d aff...

Страница 13: ...mode rpm time on off press to start temperature C on off orbital incubator SI500 Italiano Pannello di controllo della temperatura Pannello di controllo della velocit Connettore USB Fori di riferimento...

Страница 14: ...rico sia bilanciato Non fermare il movimento manualmente In caso di interruzione della rete elettrica al ripristino dell alimentazione l unit non riprende a funzionare In caso di interruzione dovuta a...

Страница 15: ...V F collegato a un connettore IEC320 N B il cavo di rete tipo UK protetto da un fusibile da 10 A montato sulla sommit del connettore IN CASO DI DUBBIO CONTATTARE UN ELETTRICISTA QUALIFICATO 4 2 Colleg...

Страница 16: ...Regolazione della temperatura Con il pannello di controllo superiore attivare l unit di controllo della temperatura premendo la manopola di comando Tenendo premuto il pulsante press to set premere per...

Страница 17: ...ncidenti uso non conforme negligenza errori di regolazione o di riparazione danni dovuti a installazioni adattamenti modifiche montaggio di componenti non approvati o riparazioni condotte da personale...

Страница 18: ...tor SI500 start stop mode rpm time on off press to start temperature C on off orbital incubator SI500 Panel de control de temperatura Panel de control de velocidad Conector USB Orificios localizadores...

Страница 19: ...ed el ctrica No debe utilizarse en atm sferas peligrosas ni junto con materiales peligrosos Aseg rese de que la carga est equilibrada No intente detener el movimiento de la unidad de forma manual En c...

Страница 20: ...LEVE A CABO UN ELECTRICISTA CUALIFICADO NOTA Consulte la placa de caracter sticas del equipo para asegurarse de que el enchufe y el fusible sean los adecuados para la tensi n y potencia indicadas Los...

Страница 21: ...peratura pulsando el bot n de control Mantenga pulsado el bot n press to set pulsar para establecer y gire el bot n hasta que se muestre la temperatura correcta Suelte el bot n press to set pulsar par...

Страница 22: ...uart o su agente La instalaci n de piezas no aprobadas puede afectar al rendimiento de las funciones de seguridad dise adas en el instrumento Para obtener una lista completa de las piezas que necesita...

Страница 23: ...o de la plataforma 335 x 335 mm Especificaciones t cnicas Rango de temperatura Ambiente de 5 C a 60 C Resoluci n de visualizaci n de temperatura 0 1 C Precisi n de temperatura 0 5 C Fluctuaci n de tem...

Страница 24: ...on off orbital incubator SI500 start stop mode rpm time on off press to start temperature C on off orbital incubator SI500 Temperaturbedienfeld Drehzahlbedienfeld USB Schnittstelle Positionierungsloc...

Страница 25: ...tromversorgung angeschlossen ist Betreiben Sie das Ger t nicht in einer gef hrlichen Umgebung oder einer Umgebung mit gef hrlichen Werkstoffen Stellen Sie eine ausgeglichene Last sicher Versuchen Sie...

Страница 26: ...F NUR VON EINEM QUALIFIZIERTEN ELEKTRIKER DURCHGEF HRT WERDEN HINWEIS Pr fen Sie anhand des Ger teschildes dass Stecker und Sicherungen f r die angegebene Spannungs und Stromst rke geeignet sind Die L...

Страница 27: ...en Bedienfeld dr cken Halten Sie die Taste press to set zum Einstellen dr cken gedr ckt und drehen Sie dabei den Knopf bis die richtige Temperatur angezeigt wird Lassen Sie die Taste press to set zum...

Страница 28: ...ng nicht genehmigter Teile kann die Leistung der in das Ger t integrierten Sicherheitseinrichtungen beeintr chtigen Um eine umfassende Ersatzteilliste anzufordern die Servicetechniker f r interne Repa...

Страница 29: ...0 mm Tischgr e 335 x 335 mm Technische Daten Temperaturbereich Umgebungstemperatur 5 C bis 60 C Aufl sung der Temperaturanzeige 0 1 C Temperaturgenauigkeit 0 5 C Temperaturschwankung 0 5 C bei 37 C Te...

Страница 30: ...Notes...

Страница 31: ...be sure that other equipment used in its vicinity will meet these standards and so we cannot guarantee that interference will not occur in practice Where there is a possibility that injury damage or l...

Страница 32: ...ambro Milano Italia Tel 39 0 2 98230679 Fax 39 0 2 98230211 e mail marketing bibby scientific it www bibby scientific it Bibby Scientific US Ltd 3 Terri Lane Suite 10 Burlington NJ 08016 USA Tel 800 2...

Отзывы: