Caution: Fuses are fitted in both live and
neutral lines.
Fuses:
230V - 3.15A anti surge fuse. For all other
voltages always replace with an equivalent fuse
type and rating as those being removed,
alternatively see rating plate for fuse type and
rating.
This unit is supplied with two mains leads fitted
with IEC plugs for connection to the instrument.
One lead has a U.K. 3 pin plug and the other
has a 2-pin “Shuko” plug for connection to the
mains. Choose the lead appropriate for your
electrical installation and discard the other.
Should neither lead be suitable, take the lead
with the U.K. plug and replace the plug with a
suitable alternative. This involves cutting off the
moulded plug, preparing the cable and
connecting to the rewireable plug in accordance
with its instructions.
IT IS IMPORTANT THAT THIS OPERATION
SHOULD ONLY BE UNDERTAKEN BY A
QUALIFIED ELECTRICIAN
NOTE: Refer to the equipment’s rating plate to
ensure that the plug and fusing are suitable for
the voltage and wattage stated. The wires in the
mains cable are coloured as follows:
LIVE -
BROWN
NEUTRAL - BLUE
EARTH -
GREEN/YELLOW
The appropriate mains lead should be connected
to the instrument BEFORE connection to the
mains supply. Should the mains lead need
replacement a cable of 1mm
2
of harmonised
code H05VV-F connected to an IEC320 plug
should be used. N.B. the U.K. mains lead is
protected by a 10A fuse mounted in the plug
top.
IF IN DOUBT CONSULT A QUALIFIED
ELECTRICIAN
4.2
Connect to the electricity supply - DO NOT
SWITCH ON.
4.3
Place the unit on a firm, level, non-slip
surface ensuring that there is sufficient free
space on all sides without coming into contact
with anything during use.
2
4.4
Open the door. Pull the two black sprung
release knobs either side of the cradle outwards
towards the sides of the incubator and draw the
cradle platform out of the incubator towards you.
4.5
Load the cradle with vessels, adjusting the
rubber-covered securing bars as required. Ensure
that the load is evenly balanced and does not
exceed the maximum load stated.
4.6
Gently push the platform back into the
incubator until it locks in place. Close the door.
N.B: Please ensure that the retractable
platform has clicked and locked in place
before attempting to operate the unit,
failure to do this can result in damage
to the unit.
5. Operation
5.1 Setting the speed
Using the keypad at the base of the instrument,
switch the unit ON by pressing the control knob.
Press the mode button until a red dot shows in
the display below “rpm”. The display will then
show the last set speed. Adjust the speed by
turning the control knob. The minimum speed is
30rpm and the maximum, 300rpm.
Once the correct speed is displayed press the
start/stop button to begin motion. The speed
can be adjusted without halting the unit. The
unit can be halted at any time by pressing the
start/stop button. If the door is opened while the
unit is in motion a warning alert will sound and
the motion will stop; the word “door” will also
be momentarily displayed on the screen. Close
the door to continue.
on/off
time
rpm
start/stop
mode
Time/speed LED
Red digital display
Start/stop button
Mode button