Fault Finding

In the event of operating difficulties with your Aquatron, it is 
suggested that the following basic checks are made. (Please note 
that these checks should only be carried out by suitably qualified 
personnel).

If these checks fail to identify and remedy the problem then you 
are advised to seek the help of your supplier or the Service 
Department of Bibby Scientific Ltd.

1.  Unit completely inoperative – No mains green indicator on.

 

Check the electricity supply to the Aquatron and ensure that a 
fuse has not blown at the supply fuse board. (This check 
should be conducted by a qualified electrician).

2.  Heater(s) not on – Mains green indicator light on, water 

flowing.

 

a. Check the Aquatron switches are in the correct mode, i.e:

 

  ON/OFF switch is in the ON position.

 

  CLEAN switch is in the OFF position.

 

  Check that the reservoir indicator light is NOT illuminated  
  and indicating a full collection reservoir.

 

b. Check that the fuse, located on the PCB inside the electrical  
  chamber of the cabinet, has not blown. (This check should  
  be conducted by a qualified electrician).

 

c. Check that the flow of cooling water is sufficient.

 

  A4000 requires a flow rate of at least 1 litre/min.

 

  A4000D & A8000 require a flow rate of at least 2 litres/min.

 

  If the flow rate is NOT sufficient then:

 

  Check the mains water supply is turned on.

 

  Check the mains water mini valve assembly is open.

 

  Check the water inlet filter is not blocked (See note 8).

 

  Check the flow of water is not impeded by blockages or  
  restrictions.

3.  Heater(s) not on – Water not flowing, mains green indicator 

light on.

 

Check that the thermostat has not operated. Confirm this by 
pressing the thermostat reset button located inside the main 
chamber of the cabinet.

 

If the thermostat has cut out then the cause should be 
identified at once.

 

Check that the boiler contains sufficient water.

 

Check that the plumbing connections are not leaking.

 

Check that the stop cock on the boiler level control WL48 is 
closed.

4.  Heater in second stage boiler of A4000D not on.

 

Check the setting of the boiler level control W4L2A. See the 
A4000D Assembly section of this manual.

 

Check thermostat has not operated. See notes under step 3.

5.  Water level in boiler too high – Choking of condenser.

 

Check that the flow of water into the boiler is not excessive. 
Control flow by adjusting the mains water mini valve assembly.

 

Check that the coolant water drain pipe falls away from the 
Aquatron without restrictions or U bends.

 

If level of water in the second stage boiler of model A4000D is 
too high, then:

 

Check setting of the float switch in boiler level control W4L2A. 
See A4000D Assembly section of this manual.

6.  Distillate quality low.

 

Check the condition of the boiler and clean if heavily scaled.

7.  Distillate collection reservoir floods.

 

Check the location of the reservoir water level sensor WLS 
inside the reservoir. See the Distillate Collection section of this 
manual.

 

Check that the WLS pipe connections between the glass bell 
and Aquatron are secure and air tight.

8.  Cleaning blocked water filters.

 

Your Aquatron is fitted with a built in water filter to prevent 
the ingress of particulate matter into the plumbing system.

 

The filter can be examined and cleaned if necessary by 
unscrewing the black plastic screw cap which secures the 
mains water mini valve assembly.

 

A similar filter is also incorporated on the same side of the 
cabinet for use with pre-treated water supplies. Access is via 
the black screw cap which secures the hose nozzle.

9.  Persistent tripping of the RCBO.

 

This is likely to be due to the integrity of the heater which 
should therefore be replaced.

Additional Checks for Aquatrons fed with deionised or pre-treated 
water.

a.  Boiler does not fill.

 

Check that supply of pre-treated water is switched on.

 

Check input water filter on side of cabinet is not blocked (See 
step 8).

 

Check water feed conversion kit A4000/WCK or A4000D/
WCK is correctly fitted and adjusted. See the Connection of 
Stills to Deionised / Pre-Treated Water Supplies section of  
this manual.

b.  Heater(s) not on.

 

Check water level in boiler is above the heating element.

 

Check water feed conversion kit A4000/WCK or A4000D/
WCK is correctly fitted and adjusted. See the Connection of 
Stills to Deionised / Pre-Treated Water Supplies section of  
this manual.

23

Содержание aquatron A4000

Страница 1: ...A4000 A4000D A8000 Aquatron Version 2 2 Automatic Water Stills Assembly and operating instructions...

Страница 2: ...ay be expected not to interfere with or be affected by other equipment with similar qualifications We cannot be sure that other equipment used in its vicinity will meet these standards and we cannot g...

Страница 3: ...tability of your supply consult a qualified electrician before installing your Aquatron still 3 Water Supply A cold water supply is required for coolant and feed purposes This may be from the mains su...

Страница 4: ...ssembly I A4000HK 1 Item Description Part no number 13 Solenoid Assembly A4000 SA 14 Mains Water Mini Valve Assembly 7001059 S 15 Reservoir Water Level Sensor WLS 16 Distillate Outlet Pipe Bench Mount...

Страница 5: ...of it into a beaker of hot water for approximately 20 30 seconds this will make the tubing more supple Connect the 16mm tubing to the boiler level control Item 22 drain connection below the Rotaflo st...

Страница 6: ...I WH48 1 B 16mm tubing x 1m I WH48 1 B Solenoid Condenser Thermostat Thermostat Glass Heater Boiler Level Control Funnel Rotaflo Stopcock 1 2 3 Water Flow Switch A4000 Installation Diagram Boiler Mai...

Страница 7: ...ck up into the boiler level control 11 Switch the CLEAN button off by pressing it a second time The switch light goes out 12 The heater will now warm up and eventually run at a red glow If the flow ra...

Страница 8: ...Water Mini Valve Assembly 7001059 S Thermostat Glass Tube WTT48 Thermostat WT4 Boiler Level Control WL48 Boiler Plastic Coupling WBC1 Heater Assembly A4000 HA Heater Assembly 220V A4000 HA220 Boiler...

Страница 9: ...SA 13 Mains Water Inlet Valve Assembly 7001059 S Item Description Catalogue no number 14 Reservoir Water Level Sensor WLS 15 Distillate Outlet Pipe 16 Drain Pipe from W4L2A 17 Cooling Water Pipe to Dr...

Страница 10: ...m x 300mm tubing connected to the rear boiler level control side arm to the distillate outlet of the front condenser Fig 2c 19 Take the front Boiler Level Control WL48 remove the end split nut and cen...

Страница 11: ...e front boiler level control drain connection below the Rotaflo stopcock Secure with a tie wrap 21 Take the 1m length of 16mm hose from hose kit I WH48 3 B Place one end into a beaker of hot water for...

Страница 12: ...rolling water and electricity inside the still The switch illuminates when pressed in the ON position Amber Light When illuminated this indicates that the distillate collection reservoir is full as de...

Страница 13: ...element if the flow is insufficient Thermostat Cut Out In normal operation the water within the glass will be at 100 C In single fault conditions i e in the event of water supply failure the content...

Страница 14: ...enoid Assembly A4000 SA Item Description Catalogue no number 14 Mains Water Inlet Valve Assembly 7001059 S 15 Reservoir Water Level Sensor WLS 16 Distillate Outlet Pipe Bench Mounted 17 Cooling Water...

Страница 15: ...et faces the front and the two water connections are parallel to the unit When slight resistance is felt when fitting the condenser press down slightly on the top of the condenser until a gap of appro...

Страница 16: ...Fit the funnel into the top of the boiler level control and tighten the black cap 20 Connect the tubing connector number 5 to the boiler level control Fig 3b 21 Take the remaining length of 8mm tubing...

Страница 17: ...l White Clean Light This shuts off the heating element but allows water to flow into the boiler when the still is being commissioned or cleaned When pressed in the CLEAN position the switch illuminate...

Страница 18: ...the reservoir is full to allow the heating element to cool It is recommended that the operation of the flow switch and reservoir control is checked on a regular basis a Simulation of Mains Water Suppl...

Страница 19: ...Bibby Scientific offer a purpose designed reservoir system for the collection of distilled water from your Aquatron WR20 A 20 litre Pyrex glass reservoir complete with Rotaflo stop cock and special l...

Страница 20: ...ionised pre treated water solenoid and secure with a jubilee clip 10 Connect the other end of the tubing assembly I WCK 02 to the boiler level control 11 Thread the earth cable from tubing assembly I...

Страница 21: ...anel 15 Refit the perspex viewing screen cabinet side panel and cabinet roof Operation 1 Before switching on either the mains electricity or water supply identify the control switches and indicator li...

Страница 22: ...the ON OFF switch to the ON position the boiler will refill to its operating level 8 Push the ON OFF switch to the OFF position 9 Allow the chemical reaction within the boiler to continue until all t...

Страница 23: ...level control W4L2A See the A4000D Assembly section of this manual Check thermostat has not operated See notes under step 3 5 Water level in boiler too high Choking of condenser Check that the flow o...

Страница 24: ...Circuit Diagram for A4000 AQUATRON Water Still A4000 PCB S T 125 m AL S WHEN USED WITH WCK N 1 4 WAY SOCKET S 1N4007 RL4 E S 12 WAY SOCKET NOTE SKT1 IS USED IN CONJUNCTION WITH THE DE IONISED FEED CO...

Страница 25: ...E SWITCH S 1 2 3 5 4 1 2 4 3 WHEN USED WITH WCK N Circuit diagram for A4000D Wiring diagram for A4000D NOTE SKT1 IS USED IN CONJUNCTION WITH THE DE IONISED FEED CONVERSION KIT WCK N IF THE STILL IS NO...

Страница 26: ...2 3 4 5 Circuit Diagram for A8000 AQUATRON Water Still A4000 PCB S T 125 m AL S WHEN USED WITH WCK N 1 4 WAY SOCKET S 12 WAY SOCKET RL3 E S 1N4007 10 11 12 7 9 3 1 6 5 2 3 2 4 1 4 WAY PLUG 12 WAY PLU...

Страница 27: ...27 Bibby Scientific Ltd Stone Staffs ST15 0SA UK Tel 44 0 1785 812121 Fax 44 0 1785 813748...

Страница 28: ...ific France B timent le Deltaparc Silic pars Nord 2 7 rue du Canal BP 55437 Villepinte 95944 ROISSY Charles de Gaulle France Tel 33 0 1 48 63 78 00 Fax 33 0 1 48 63 78 01 Email contact bibby scientifi...

Отзывы: