25
Care and Maintenance
Inspections by an Authorized Service Technician
Inspections for lift
cassette and track
system (continued)
Initially
Before
every use
Every two
months or
500 cycles
Every four
months or
1000
cycles
Every
year or
2500
cycles
Every two
years or
5000 cycles
Clean the rail and the clip-on
charging station contacts.
X
Overall inspection by
authorized personnel.
X
Verify emergency stop cord.
X
Verify emergency lowering
device.
X
FREQUENCY
Inspections for
spreader bar and
slings
Initially
Before
every use
Every two
months or
500 cycles
Every four
months or
1000 cycles
Every
year or
2500
cycles
Every two
years or
5000 cycles
Inspect all sling parts
(attachments, fabric, stitch
areas and strap) for signs of
wear, discoloration,
deterioration or loose threads.
X
Clean sling as indicated on the
tag.
When necessary
Inspect the spreader bar on the
strap of the lift for damage or
cracks. Make sure all
attachments are properly
secured (e.g. split ring).
X
X
FREQUENCY
Inspection for lift
cassette
Initially
Before
every use
Every two
months or
500 cycles
Every four
months or
1000
cycles
Every
year or
2500
cycles
Every two
years or
5000 cycles
Replace strap.
X
Inspect frame parts interlock
and hardware for malfunction
and make sure there are no
parts missing.
X
Inspect gears for wear.
X
FREQUENCY