MAINTENANCE
Portable VOYAGER rev. 6
15
Replace the fuse with the following:
Fuse: GMC 0.5 A 250 V North America (When
input is 120 Vac) Product # C8FGMC0.5
GMC 0.25 A 250 V European (When input
is 240 Vac) Product # C8FGMC0.250
The fuse may also be ordered through your local
dealer or representative.
6.7 Sling Inspection
•
Confirm the sling is a Sunrise Medical sling.
•
Check the load bearing straps for wear or
fraying
•
Check the sling straps are securely stitched
•
Check the body of the sling for wear, cuts in
the fabric.
•
Check any sling accessories (ie additional
spreader bars / side suspenders) for wear on
the hooks or central suspension point.
6.8 Annual Inspection
The Voyager Portable and its accessories must
be inspected annually by a certified technician.
VOYAGER
PORTABLE
AND
ACCESSORIES
MUST
BE
SERVICED EVERY 12 MONTHS AS
A MINIMUM REQUIREMENT. IN THE
UK THE VOYAGER PORTABLE
MUST
BE
INSPECTED
AND
MAINTAINED
IN
COMPLAINCE
WITH 1998 No 2307 HEALTH &
SAFETY: LIFTING OPERATIONS
AND
LIFTING
EQUIPMENT
REGULATIONS 1998.
Do not attempt to do the inspection unless
you are certified to do so.
As part of the annual inspection, an annual
load test with the safe working load must be
performed on the VOYAGER PORTABLE as
required for CE MARK MAINTENANCE.
After annual inspection, have the log book
signed by the certified technician.
6.9 UK Regulations
IN THE UK THE VOYAGER PORTABLE
MUST BE INSPECTED AND MAINTAINED IN
COMPLAINCE WITH 1998 NO 2307 HEALTH
& SAFETY: LIFTING OPERATIONS AND
LIFTING EQUIPMENT REGULATIONS 1998.
YOUR DEALER WILL PROVIDE YOU WITH
MORE INFORMATION