MAINTENANCE 

Portable VOYAGER rev. 6 

14 

Loose threads in stitched area.

Noticeable discoloration by having

a lighter color strap than the double

Side wear.

Middle wear.

thickness stitched area.

 
 

If there are any signs of wear as 
indicated  here  or  other  visual 
defects, 

strap 

should 

be 

changed 

immediately. 

 

By 

continuing to use the lift without 
changing  the  strap,  caregiver 
and  patient  safety  is  greatly 
compromised. 

 
In 

any 

case, 

manufacturer 

recommends  to  change  strap  at 
least  every  two  years.    By 
continuing to use the lift without 
changing  the  strap,  caregiver 
and  patient  safety  is  greatly 
compromised. 
 
Take 

note 

of 

results 

of 

inspection in the log book at the 
end of this manual. 

 

6.4  Handling and Storage 

 
Avoid violent shock during transportation. 

 

The lift should not remain stored for long periods 
of time without recharging the batteries. 
 

BHM 

Medical 

recommends 

charging  of  batteries  at  least 
every two weeks even if the lift is 
not  used.    This  will  prevent 
premature aging of batteries. 

 

6.5  Battery Replacement 

BHM  uses  sealed  lead-acid  batteries.  Contrary 
to  nickel-cadmium,  BHM  batteries  do  not  have 

any  memory  effect.  Therefore,  batteries  should 
not be completely drained before recharge. 

Replace the battery when it has been dropped or 
if there is a noticeable reduction in the number of 
transfers  that  can  be  performed  between 
charges.    If  you  notice  the  Voyager  Portable 
“beeping’ 

– 

see 

the 

instructions 

in 

“Troubleshooting”  section  to  ensure  it  is  a 
problem with the battery. 

To replace a battery, contact your local dealer or 
representative and order product # A8500. 

DO  NOT  ATTEMPT  TO  USE  A 
BATTERY NOT AUTHORIZED BY 
BHM MEDICAL. 
 
BHM  batteries  are  specially 
designed  for  BHM  charging 
systems.    Attempting  to  use  an 
unauthorized 

battery 

may 

seriously  damage  the  lift  and/or 
the charger. 

 

6.6  Charger Fuse Replacement 

(Charger #A8200 only)

 

 
If  the  charger  light  does  not  illuminate  when 
there is a battery in it, do the following steps: 

 

1.  Make  sure  the  battery  is  properly  placed  in 

the charger.  The silver contact points should 
be inside the charger. 

2.  Check  to  make  sure  the  power  cord  is 

correctly  plugged  in  to  the  charger  and  the 
wall.  

3.  If the light still doesn’t work, plug the charger 

into  a  different  outlet  in  a  different  area  to 
ensure that the outlet is working properly. 

4.  Pull  the  power  cord  out  of  the  charger.  

Under  the  socket  on  the charger,  there  is  a 
drawer  or  magazine  where  there  is  a  fuse.  
Using  a  pen or  screwdriver,  pull  the drawer 
out  gently.    Pull  the  burnt  fuse  out  of  the 
drawer  that  is  next  to  the  charger  (see 
diagram  below)  and  discard.    Take  the 
replacement  fuse  and  place  it  in  the  same 
slot  where  the  burnt  fuse  was.    Slide  the 
drawer  back  into  the  charger.    Plug  the 
power  cord  into  the  charger.    Place  the 
battery  into  the  charger  and  check  for  the 
light. 

5.  If  the  light  does  not  illuminate,  contact  your 

local dealer or representative for assistance. 

PDF created with pdfFactory Pro trial version 

www.pdffactory.com

Содержание OXFORD VOYAGER PORTABLE

Страница 1: ...vicing this system VOYAGER PORTABLE IMPORTANT Read these instructions before installing operating or servicing this system USER MANUAL Revision 4 Jan 2002 1 1 1 DO NOT DESTROY KEEP KEY ENCLOSED AND MA...

Страница 2: ...ry C Bath Category D Toilet shower chair Category E Floor Category F 90 degree rotation The Voyager series are suitable for patients in the sitting sitting recumbent and recumbent positions Approvals...

Страница 3: ...OW TO USE THE VOYAGER PORTABLE 8 4 2 EMERGENCY LOWERING FEATURE 9 4 3 SLING USAGE 10 5 BATTERY INFORMATION 10 5 1 CHARGER SPECIFICATIONS A8200 11 5 2 CHARGING THE BATTERY WITH CHARGER A8200 11 5 3 CHA...

Страница 4: ...nual product number and equipment identification numbers 1 2 Equipment Identification The unit s identification number specification model serial number appears on a silver nameplate attached to the b...

Страница 5: ...technician for servicing IMPORTANT Keep all components of the lift clean and dry and have electrical and mechanical safety checkpoints done as instructed in the Maintenance section of this manual Repl...

Страница 6: ...or store the battery under direct sunlight or near a heat source Do not dispose of in fire Do not short the battery terminals Do not incinerate Flush with water if electrolyte Acid comes in contact wi...

Страница 7: ...as a car safety belt mechanism Quick Release feature allows very fast winding and unwinding of the strap before and after transfer operation Anti crush feature stops the lift when there is no tension...

Страница 8: ...on the strap Make sure the strap has some slack in it before attempting to slide the knob over DO NOT force the knob It should slide easily 3 Attach the round hook to the trolley in the track To help...

Страница 9: ...ure the strap has some slack in it before attempting to slide the knob over DO NOT force the knob It should slide easily When using the QUICK RELEASE hold the strap in your hand and release the knob 1...

Страница 10: ...sfers per battery charge may vary depending on patient weight and battery condition See graph below Life cycle number of charging cycles of the battery is largely dependent on the depth of discharge i...

Страница 11: ...tep instructions Eject the battery from the lift unit by using the Battery Eject Button 1 Keep your hand near the opening to catch the battery 2 Keep one hand near the opening for the battery 2 before...

Страница 12: ...n be left in the charger for an extended period of time without damaging either the charger or battery If the green light does not go on when the charger is plugged into an electrical outlet see Troub...

Страница 13: ...her damage If there is any evidence of damage DO NOT USE IT Contact your local representative to have the lift serviced Inspect the sling for tears frayed straps or loose stitching If the sling has an...

Страница 14: ...Portable beeping see the instructions in Troubleshooting section to ensure it is a problem with the battery To replace a battery contact your local dealer or representative and order product A8500 DO...

Страница 15: ...ted annually by a certified technician VOYAGER PORTABLE AND ACCESSORIES MUST BE SERVICED EVERY 12 MONTHS AS A MINIMUM REQUIREMENT IN THE UK THE VOYAGER PORTABLE MUST BE INSPECTED AND MAINTAINED IN COM...

Страница 16: ...nspect for any defect or loose threads in the stitched areas X Cleaning sling as indicated on the tag When Necessary ACCESSORIES Verify strap extension for wear X Verify arm extension for wear X MAINT...

Страница 17: ...have been checked without success you probably have a motor or electric problem you then need to call your BHM Medical representative The unit starts and stops repetitively If the load is more than s...

Страница 18: ...e inside the charger Contact your local dealer or representative The battery does not slide into the lift Check if there is any damage to the battery that would interfere in the lift battery s slot IF...

Страница 19: ...duct No C8FGMC0 5 120 24 volt North America C8FGMC0 25 240 24 volt European HANDSET Product No 700 13600 HOOK Product No Q8000 STANDARD TROLLEY Product No A5900 SWIVEL TROLLEY Product No A5910 ARM EXT...

Страница 20: ...repairs required Withdraw the patient lifter from service if inspection reveals that user safety is jeopardized in any way from use of the patient lifter TO BE COMPLETED AFTER EACH SERVICE OR INSPECT...

Страница 21: ...___________ Inspected by ____________________ ____________________ Printed Signature Service Type Pre delivery Periodic inspection Minor Major Condition report Action taken Date _________________ Insp...

Страница 22: ...___________ Inspected by ____________________ ____________________ Printed Signature Service Type Pre delivery Periodic inspection Minor Major Condition report Action taken Date _________________ Insp...

Страница 23: ...sent its original proof of purchase at the moment of the claim The defective unit assembly or part must be returned to BHM Medical for inspection The part or components repaired or replaced are guaran...

Отзывы: