The pushing 

ram cannot 

be pulled out

The portion of the pushing ram that 

enters the interior of the heating 

chamber is fixed by the cooling of the 

gutta-percha.

1. Power on and set the 

temperature to 200 °C. After the 

temperature reaches the set value, 

pull out the pushing ram back;

2. Contact your local dealer or our 

company.

Charging

failure after

connecting to 

power supply.

1. The power supply is not

correctly connected;

2. The power supply is

damaged, or the specification

doesn’t match.

3. There are impurities on the

contact thimble of charging

base.

1. Unplug and reconnect.

2. Replace the battery.

3. Wipe the thimble with alcohol, 

dry it, and reconnect.

The service

time after 

each

charging is

shortened.

The battery capacity becomes 

smaller.

Send to the repair center.

ERROR code 

appears 

on display 

screen

The heating chamber is damaged.

1. Send to the repair center.

2. Contact local distributor or 

manufacturer.

If the problem still cannot be solved, please contact your local dealer or Beyes Dental Canada 

Inc.

1. When the pushing ram is in the the GPro F1 gutta percha obturation gun, please do not 

push or pull the pushing ram vigorously. When the gutta-percha is heated up to the preset 

temperature, the pushing ram should be pushed by pulling the trigger for multiple times. If the 

pushing ram does not move, try to push it manually with a slight force, and try to pull the trigger.

2. Please refer to the recommended temperature to set the preset temperature.

3. To remove all remaining material, please first remove the injection needle, and then pull the 

trigger to squeeze out all the residual material in the heating chambers. Be careful not to touch 

the head of the the GPro F1 gutta percha obturation gun to avoid scalding. Power off, cool it 

down slightly, and push the pushing ram down.

10. After-sales service

Starting from the date of sales, if the device does not function up to standard due to 

manufacturer defect, our company will be responsible for the repair of device during its warranty 

period. Please refer to the warranty card for warranty period and scope.

Содержание GPRO F1

Страница 1: ...Gutta Percha Obturation System Instruction for Use...

Страница 2: ...ending on the level of risk involved safety requirements are classed under the following indications Danger Always refer to personal injury Warning Refer to possible damage to property Table of Conten...

Страница 3: ...traindications Contraindications to the use of this device on patient includes Patient allergic to known natural latex and metals such as stainless steel silver copper Patient with hemophilia Patient...

Страница 4: ...attery cover 10 Charging indicator 11 Charging base 12 Power adapter plug 13 Power adapter 14 Connecting hole for power supply 15 Trigger 16 Ram release 1 5 Diagram of Device The Gpro F1 is equipped w...

Страница 5: ...not drain battery Always keep battery fully charged to maintain battery life Recharge battery if the device has not been used for more than 1 month 2 2 Install needle Warning Make sure the device is o...

Страница 6: ...Gutta percha Select the right size of gutta percha for operation Insert gutta percha into the loading slot Warning Slot fits only 1 gutta percha pellet Too many pellets can damage the device Use ram...

Страница 7: ...gutta percha Load gutta precha pellet see installation Make sure the temperature reaches the set temperature Squeeze and release the trigger The ram then steps to push the gutta percha through the nee...

Страница 8: ...the trigger to extrude the melted gutta percha obturator into the canal 4 Slowly retract the needle with continuous flow while withdrawing the tip from the canal 5 Follow with use of GPro P1 heating p...

Страница 9: ...power adapter Optional preset temperatures 150 C 180 C 200 C 230 C Time consumption for charging About 2 5h First charging needs 3 h Power supply Input AC100V 240V 50 60Hz 800mA Output DC15V 1 6A Batt...

Страница 10: ...tch Ordinary equipment Used indoor only Device complies with WEEE directive Can be autoclaved Keep dry Handle with care Recovery Attention Please refer to the accompanying documents Humidity limit for...

Страница 11: ...tly installed to prevent from falling off or gutta percha leakage during operation 10 Do not use excessive force when pre bending the injection needle to prevent the injection needle from breaking Whe...

Страница 12: ...ice is not used please turn off the power and unplug the power supply plug If the the GPro F1 gutta percha obturation gun is in a low battery state for a long time the service life of battery will be...

Страница 13: ...nnot be solved please contact your local dealer or Beyes Dental Canada Inc 1 When the pushing ram is in the the GPro F1 gutta percha obturation gun please do not push or pull the pushing ram vigorousl...

Страница 14: ...ty period measured from the date of invoice to the original user shall be as follows GPro F1 are warranted for a period of 12 months 11 3 EXCLUSIONS This limited warranty does not cover and BEYES shal...

Страница 15: ...r delays including but not limited to damages for loss of profits or income loss of use downtime cover and employee or independent contractor wages payments and benefits 11 5 WARRANTY DISCLAIMER This...

Страница 16: ...aplein 10 Ie Verd 2595AA The Hague Netherlands Tel 31645171879 English 31626669008 Dutch Email info beyes ca Website www beyes ca Printed in Canada ENI035 Rev 1 11 03 21 Federal law restricts this dev...

Отзывы: