Beurer GL30 mg/dL
28
EMC
This monitor conforms to European Standard EN 61326 and is subject to particular pre-
cautions with regard to electromagnetic compatibility (EMC). Please note that portable
and mobile HF communication systems may interfere with this instrument. For more
details, please contact customer service at the address indicated.
Function of the test strips
The test strips provide a quantitative measurement of the glucose content of the capillary
whole blood. When the slot for blood taking comes into contact with a drop of blood, it is
automatically filled by simple capillary action. The blood is drawn into the absorption slot
in the test strip and the monitor measures the glucose content of the blood.
The test is based on the measurement of an electric current that is generated by the
chemical reaction of the glucose with the reagent on the strip. The monitor analyses the
current. The current depends on the glucose content of the blood sample. The results are
shown on the display of the blood glucose monitor. Only a very small volume of blood is
required (0.5 microlitre) and the actual measurement takes about six seconds. The test
strips can measure glucose values from 20 to 600 mg/dL.
Chemical constituents of the test strip sensor
•
Glucose oxidase 10%
•
Electron shuttle 50%
•
Enzyme protection 8%
•
Non-reactive constituents 32%
Function of the control solution
The control solution contains a specific proportion of glucose, which reacts with the test
strip. A test with control solution is similar to a blood test. The difference is that a drop of
control solution is used instead of a drop of blood. The test result with the control solution
must be within the result range. The result range is printed on every test strip container.
Chemical composition of the control solution
The control solution is a red solution with a D-glucose content of less than 0.2%.
Ingredients
Percentage
D-glucose
0.05 – 0.19%
Salts
1.4%
Substance for regulating viscosity
15.0%
Controls
The
Beurer GL30 mg/dL blood glucose monitoring system
for self-testing conforms to
the following European Directives and standards: IVD (98/79/EC), EN 61010-1,
EN 61010-2-101, EN 13640, EN ISO 14971, EN ISO 15197, MDD (93/42/EC).