The  control  solution  is  used  to  test  the  complete  blood  glucose  monitoring  system.  It  is 
used to check that the monitor and test strips operate together correctly and whether the 
test is conducted correctly. You should perform a control solution test if you suspect that 
the measuring device or test strips may be faulty, or if you repeatedly receive unexpected 
blood sugar results. Also test the measuring device if it has been dropped or damaged. 
The  control  solution  is  available  separately.  When  performing  the  control  solution  test, 
please follow the additional instructions in the control solution user guide. 
 

 

CAUTION

 

Never  use  control  solution  from  other  manufacturers.  It  will  be  impossible  to 
check  that  your  Beurer  blood  glucose  monitor  operates  correctly  with  other 
control solutions. 

 
Testing with control solution 

Hold the monitor with the display facing you.  

Insert  a  test  strip  into  the  monitor  with  the  contacts  first.  Make  sure  that  the 
front of the test strip is facing you (see "3.3 Test strips" page 11). 

The  monitor  switches  on  automatically  and  shows  consecutively  the  initial 
display, the ambient temperature, and the code number. The monitor is ready 
for a test as soon as the flashing blood drop symbol 

6

 is displayed.  

Check the code number. If the numbers are not identical, remove the test strip 
and code the monitor (see "4.3 Checking code number (coding)" page 14). 

Press the M button to switch to control mode. In control mode test results are 
not saved and your statistics are not skewed.  

“CTL” is shown on the display. 

Shake the control solution well before use. Unscrew the cap and squeeze out 
one drop. Wipe the first drop away and squeeze out another drop.  

Do not apply the drop directly to the test strip, otherwise the control solution 
remaining  in  the  vial  tip  may  be  contaminated  by  contact  with  the  test  strip. 
First place the drop on your fingertip or on a clean surface. Then hold the drop 
to  the  slot  for  taking  blood  on  the  test  strip.  The  solution  is  sucked  into  the 
slot. Wipe the tip of the vial with a clean, dry paper towel. 

When the slot is filled with solution, the monitor carries out the test. The moni-
tor  counts  back  for  approximately  six  seconds.  The  result  is  shown  in  the 
display.  

10 

Check that the test result is within the specified range of the control solution. 
The result range is printed on the test strip container. 

 

Содержание GL30 mg/dL

Страница 1: ...BLOOD GLUCOSE MONITORING SYSTEM Step by step instructions for use GL30 GB Beurer GmbH S flinger Stra e 218 89077 Ulm Germany www beurer medical com...

Страница 2: ......

Страница 3: ...or 8 3 3 Test strips 11 4 Operation and Basic Settings 13 4 1 Inserting and replacing batteries 13 4 2 Making basic settings 13 4 3 Checking code number coding 14 5 The measurement 15 5 1 Taking a blo...

Страница 4: ...itor GL30 mg dL REF 162 544 B 10 test strips REF 162 511 C 1 lancing device with AST cap for taking blood samples at different parts of the body REF 162 501 D 10 sterile lancets REF 162 513 E 1 handy...

Страница 5: ...date and time setting the temperature unit C or F coding the monitor The monitor also permits the following control functions display of the ambient temperature and warning of unsuitable temperatures...

Страница 6: ...countries Not for reuse single use only Green Dot German dual waste disposal system Use by Contents sufficient for n tests Maximum storage period in months after opening Art Nr Order number Batch Ster...

Страница 7: ...rposes they are not a substitute for a medical examination Discuss your readings with a doctor regularly Never change your doctor s instructions without dis cussing your changes with the doctor Accura...

Страница 8: ...ritation if inhaled or swallowed Keep the container away from small children Batteries saving readings WARNING Batteries can be fatal if swallowed You should store the batteries out of reach of small...

Страница 9: ...ly to prevent injury and infection of other persons Dispose of test strips and lancets in a sharps container after use Note Used completely discharged batteries must be disposed of in specially marked...

Страница 10: ...lancets 1 2 3 4 5 6 1 Cap 2 Safety shield of lancet 3 Sterile lancet 4 Lancet holder 5 Trigger button 6 Trigger setting mechanism 3 2 Blood glucose monitor Front Back 1 Slot for test strips 2 LCD dis...

Страница 11: ...est strips 3 Results display 4 Low blood glucose low 5 High blood glucose high 6 Test strip symbol 7 Speaker non functional 8 Blood drop symbol 6 9 Blood glucose unit mg dL 10 Blood glucose unit mmol...

Страница 12: ...ay 16 Time hour minute 17 Replace battery symbol 18 CTL for control solution 19 DAY AVG for average reading 7 14 21 28 60 90 days 20 AM morning 21 PM afternoon Note The blood glucose monitor is suppli...

Страница 13: ...trip into the unit with the contacts first Make sure that the front of the test strip is facing you Read carefully the following information on the handling and storage of your test strips All instruc...

Страница 14: ...onths on the la bel The storage time is reduced if it extends beyond the use by date see date beside the hourglass icon Do not use the test strips if one of the use by dates has expired The test strip...

Страница 15: ...t takes longer than usual see 4 2 Making basic settings page 13 3 Insert two new AAA 1 5 V alkaline batteries Make absolutely sure that you insert them with the correct polarity as marked Do not use r...

Страница 16: ...ayed and the Memory symbol starts blinking Caution all previously stored results will be deleted if you press M now Press the SET button OFF is displayed briefly and the moni tor automatically switche...

Страница 17: ...od sample Preparing to take the blood sample 1 Select part of body for blood sample You can take blood samples with the lancing device from the fingerprints or other parts of the body such as palm of...

Страница 18: ...e other hand Repeated puncture wounds at one puncture site may cause inflammation loss of feeling or scarring Do not use the AST cap for taking a blood sample from the finger Never compress the finger...

Страница 19: ...at the blood remains in drop form and is not smeared Blood sample from the fingerprint Position the lancing device firmly and slightly off centre on the fingerprint Press the trigger button Remove the...

Страница 20: ...contact with them This will pre vent injury and infection of other people 12 Screw the cap grey on again 5 2 Measuring blood glucose 1 Hold the monitor with the display facing you 2 Insert a test stri...

Страница 21: ...age 25 7 Remove the test strip from the monitor Dispose of the test strip as directed to prevent infection of other persons Note Do not apply more blood if the monitor does not start the measurement R...

Страница 22: ...g dL Regular self monitoring high 121 mg dL and above If this value still remains high two hours after your last meal it may indicate hyperglycaemia high blood sugar Talk to your doctor about any requ...

Страница 23: ...atically and shows consecutively the initial display the ambient temperature and the code number The monitor is ready for a test as soon as the flashing blood drop symbol 6 is displayed 4 Check the co...

Страница 24: ...You did not dispose of the first drop of control solution The top of the vial was not wiped clean The vial was not shaken sufficiently Correct the cause and repeat the test Control solution has expire...

Страница 25: ...lues You can display the average values of your blood glucose readings for the last 7 14 21 28 60 and 90 days At the same time the monitor shows how frequently you have con ducted tests over this peri...

Страница 26: ...seconds The display briefly reads OFF 5 The monitor switches off automatically 6 2 Deleting stored values 1 The monitor must be switched off Remove the battery cover from the back of the monitor 2 Pr...

Страница 27: ...d protect it against any impact You should protect the monitor from moisture dirt dust blood control solution or water major temperature fluctuations and direct sunlight Do not use the monitor in the...

Страница 28: ...es not start after inserting the test strip into the monitor and applying the blood Cause Remedy Blood volume too low Repeat test with new test strip and greater volume of blood Faulty test strip Repe...

Страница 29: ...ys Automatic shutoff Three minutes after last activity Storage and transport temperature Temperature 4 C 40 C Relative humidity 10 85 Operating ranges Temperature 10 C 40 C Relative humidity 10 85 non...

Страница 30: ...lay of the blood glucose monitor Only a very small volume of blood is required 0 5 microlitre and the actual measurement takes about six seconds The test strips can measure glucose values from 20 to 6...

Страница 31: ...ntion This guarantee does not affect your statutory rights In order to make a claim within the warranty period the customer is required to provide proof of purchase Claims must be made within a period...

Страница 32: ...GL30_mg dL_0309_GB Subject to error and change...

Отзывы: