![Beurer GL30 mg/dL Скачать руководство пользователя страница 24](http://html1.mh-extra.com/html/beurer/gl30-mg-dl/gl30-mg-dl_instructions-for-use-manual_2743989024.webp)
Beurer GL30 mg/dL
22
Expected results
At room temperature the test results with the control solution should be within the range
printed on the test strip container for approximately 95% of all tests.
WARNING
The results range printed on the test strip container is only applicable to the
control solution.
It is not a recommended value for your blood glucose
level.
If the results are outside the specified range, check the following possible causes:
Cause
Measure
The code number on the display is not
the same as that of the test strip con-
tainer.
Code the monitor (see "4.3 Checking
code number (coding)" page 14).
•
You did not dispose of the first drop of
control solution
•
The top of the vial was not wiped
clean.
•
The vial was not shaken sufficiently.
Correct the cause and repeat the test.
Control solution has expired or is con-
taminated.
Repeat the test with a new vial of control
solution.
Control solution or monitor is too warm or
too cold.
Bring control solution and monitor to
room temperature (+20°C to +25°C) and
repeat test.
Damaged test strips
Repeat the test with a new test strip.
Old test strips
Open a new test strip container. Repeat
the test.
A problem with the monitor
Contact our Customer Service depart-
ment.
WARNING
If you receive repeated results outside the specified range with the control
solution,
do not continue using the system to measure your blood glu-
cose level
. Contact our Customer Service department.