33
EMC Compliance Declaration
ELECTROMAGNETIC COMPATIBILITY INFORMATION
Table 1
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture’s declaration – electromagnetic emissions
The BF 525 is intended for use in the electromagnetic environment specified below.
The customer or the user of the BF 525 should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The BF 525 uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely to cause any inter-
ference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
Table 2
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer’s declaration – electromagnetic immunity
The BF 525 is intended for use in the electromagnetic environment specified below. The customer or the user
of the BF 525 should assure that it is used in such an environment.
IMMUNITY test IEC 60601 test level Compliance level Electromagnetic environment – guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30 %.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a typi-
cal commercial or hospital environment.
Содержание BF 525
Страница 36: ...36 760 73_BF525_2018 02 21_01_IM_BEU USA ...