84
16. Technical data
Model No.
IPO 61
Type
PO 60
Measurement method Non-invasive measurement of arterial oxygen saturation of
haemoglobin and pulse rate in finger
Measurement range
SpO₂ 0 – 100%, pulse 0 – 254 beats/minute
Accuracy
SpO₂ 70 – 100%, ± 2%, pulse 30 – 250 bpm, ± 2 beats/minute
Dimensions
L 58.5 mm x W 31 mm x H 32 mm
Weight
Approx. 27 g (without batteries)
Sensor to measure
SpO₂
Red light (wave length 660 nm); infra-red (wave length 905
nm); silicon receiver diode
Permissible operating
conditions
+10°C to +40°C, ≤ 75% relative humidity, 700 – 1060 hPa
ambient pressure
Permissible storage
conditions
-40°C to +60°C, ≤ 95 % relative humidity, 500 – 1060 hPa
ambient pressure