43
WWW.BERTIN-INSTRUMENTS.COM
USER MANUAL
”communication module” of the
DataVIEW
or
DataEXPERT
software recognizes
the set baud rate of the
AlphaGUARD
automatically.
The transfer of a several days lasting measuring series out of the value storage of
the
AlphaGUARD
into the working memory of a PC lasts with 115 200 Bd some
seconds, max. a few minutes depending on amount, operation mode, etc. With
appreciably lower baud rates this process can take several minutes more. The set
baud rate can be changed by selecting the required rate and pressing the Enter-
key for acknowledgement.
Set RS232baudrate
1 1 . 0 8 . 1 6 1 2 : 2 7 : 4 7 4 m i n
1 2 0 0 B d
3 0 0 B d
4 8 0 0 B d
1 1 5 2 0 0 B d
3 8 4 0 0 B d
U s e ↑ ↓ t o c h a n g e a n d
p r e s s < ┘k e y t o a c c e p t !
9 6 0 0 B d
1 9 2 0 0 B d
Figure 45: Selecting the baud rate for the RS232 interface
Setting of the baud rate can also be carried out using the DataVIEW or
DataEXPERT communication module via PC.
Note
In order to avoid disturbances with the AlphaGUARD monitor, if possible please
use the delivered RS-232/USB cable. Longer cables can cause disturbances
due to EMI interferences and thus malfunctions of the instrument. For on-line
operation over longer distances with RS-232 cable the AlphaGUARD interface
must be galvanically separated from the cable connection.
Important
Only use interface modules with an own power supply.
Note
Содержание ALPHAGUARD D2000
Страница 12: ...12 WWW BERTIN INSTRUMENTS COM USER MANUAL Figure 3 AlphaGUARD front side Figure 4 AlphaGUARD handle side...
Страница 82: ...82 WWW BERTIN INSTRUMENTS COM USER MANUAL Figure 65 Range of delivery 3 1 10 11 5 4 7 2 6 8 9 12 13...
Страница 93: ...93 WWW BERTIN INSTRUMENTS COM USER MANUAL Flow diagrams of the AlphaGUARD menu 14 3 1 Acquisition windows...
Страница 113: ...113 WWW BERTIN INSTRUMENTS COM USER MANUAL...